background image

19

II Chest measurement

If palm measurement is not ideal, please adopt chest 

measurement. Hold the device with the right hand. Make 

sure that the palm and fingers touch the metal electrode 

I and II reliably. Place the electrode III on bare skin about 

5cm away below the left nipple. The ECG signal detected 

by Chest measurement is similar to Lead V ECG signal.

Improper operating method

A. The contact place of electrode III deviates from left chest.

B. The holding gesture is incorrect.

C. The electrode can not contact bare skin completely.

(For example place the electrode on clothing.)

III Leg measurement

If palm measurement is not ideal and it is not convenient to use chest 

measurement, the leg measurement is recommended. When measuring, hold 

the device with the right hand. Make sure that the palm and fingers touch the 

metal electrode I and II reliably. Place the electrode III on the skin 10cm away 

above the left foot ankle. The ECG signal detected by Leg measurement is 

equivalent to Lead II ECG signal.

Improper operating method

A. You moved your foot.

B. Wrong holding gesture.

C. The electrode III can not contact bare skin completely (For example, place 

the electrode III on the sleeve of trousers or socks.).

IV Lead wire measurement

For the sake of obtaining high quality ECG, lead wire measurement can 

be used. Connect the lead and the device effectively by lead wire socket. 

Place the electrodes and connect the lead wires as the figure below to 

obtain the Lead II ECG signal; if you want to measure Lead I and Lead III 

ECG signal, please connect the lead wires to the electrodes according to 

the following table.

Improper operating method

A. Wrong in connecting lead port.

B. Place the electrode in the wrong part of your body.

ECG Leads Configuration and Electrodes Location Table

Electrode                                                Lead

Location

Lead

I

Lead

II

Lead

III

The intersection between the centerline of the right 
clavicle and Rib 2.

RA

RA

LA

The intersection between the centerline of the left 
clavicle and Rib 2.

LL

LA

RA

Between the left edge of the breast bone and Rib 5

LA

LL

LL

ENGLISH

Summary of Contents for Cardio-A Palm ECG

Page 1: ...ENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual befo...

Page 2: ...amination is suggested for normal examination of your heart condition 10 Connect Disconnect data cable to the monitor gently and carefully do NOT wring or shake the data cable plug with force to inser...

Page 3: ...prises main board LCD electrodes etc 1 4 Features Two input connections for ECG signal Embedded metal electrodes Lead wire connector optional Two output modes for ECG data LCD display Serial output in...

Page 4: ...ion the device will enter into a preparing phase this phase will last 6 seconds see the beeside picture during this period the device will judge whether the placed position is befitting Please adjust...

Page 5: ...ECG signal detected by Leg measurement is equivalent to Lead II ECG signal Improper operating method A You moved your foot B Wrong holding gesture C The electrode III can not contact bare skin comple...

Page 6: ...m is determined at the beginning of the measurement automatically 2 1 2 and 1 3 2 1 The measuring time lasts about 30 seconds Then enter the analysis phase in turn Note Refer to the period of result d...

Page 7: ...and press to measure ECG Refer to 3 2 for details 3 3 2 The function of waveform review 1 On the main menu screen press the key or of Navigation key to select Waveform Review and press to enter the re...

Page 8: ...esults and analysis Do not attempt self diagnosis 2 5 If the detected signal is too weak the beeside description can be seen Please try again 2 6 On the result description screen press the power back...

Page 9: ...he key or to move the cursor for selecting setting option and then press the key or to set numerical value 3 Press the power back button to complete the setting and return to the previous menu 3 3 5 2...

Page 10: ...ls Notes During uploading the keys and power key are effective After upload turn off the device and restart it again to start measurement Keep the monitor and data cable still during uploading data Un...

Page 11: ...tal electrode with medical alcohol when the measuring signal is too weak Please wipe the surface of the unit with clean cloth or let it dry in air Please take out the batteries if the unit will not be...

Page 12: ...etic interference 4 Weak signal Trouble ECG waveform excursion strong irrelevant waveform Solution 1 Clean skin with soap and water and eliminate skin crumb and grease to make the skin wet and greasel...

Page 13: ...rences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected products must be returned only to the d...

Page 14: ...dal Contraction Premature Ventricular Contraction or multi focus premature beat and help cure Testee s Symptom Normal heart beat is followed by a premature beat 4 Bigeminy It is a type of PVC in which...

Page 15: ...beat The display of missing beat may be doubted as Missed beats Symptom arrhythmia a long time interval Indication May be the conduction system of heart blocks it often occurs in functional cardiovas...

Reviews: