background image

17

1. 

  (Power/Back button): Long time pressing (about 2 seconds) powers on or off the device; short time pressing 

exits the operation, or back to upper level menu.

2. 

 (Navigation key): The main operating button group for this device. Press the four directions’ key for function 

selection or parameter setting; press the left navigation key for turning on/off the LCD backlight quickly.

3. Battery cover: Insert/take out the battery by opening/replacing the cover.

4. 

 (Confirmation key): Press this key for the selection at different levels of menu.

5. Metal electrode I, II: When measuring, both selectrodes should be held in the right hand.

6. Metal electrode III: When measuring, it should be held in the other hand or on the chest or on the left leg.

7. LCD screen: Display ECG waveform and different parameters.

8. Data interface port: Transmit data memories to the computer by this port (optional).

9. Lead wire socket: the lead wire connector shall be inserted into this socket if the measurement of ECG is done by 

lead wires.

10. Hanging hole: for mounting the hanging cord.

1.2 Name and Model

Name: 

Cardio-A Palm ECG

Model: 

PC-80A

1.3 Structure

The Cardio-A Palm ECG Monitor comprises main board, LCD, electrodes etc.

1.4 Features

Two input connections for ECG signal:

- Embedded metal electrodes

- Lead wire connector (optional).

Two output modes for ECG data:

- LCD display

- Serial output interface.

The features are described in detail as follows:

- Measure ECG easily at any time.

- ECG waveform and measurement results are displayed clearly on the LCD screen.

- 8 types of arrhythmia can be analyzed and presented.

- 30-second records each ECG measurement.

- 24 records can be reviewed or sent to PC by a serial port.

- Shut down automatically while no key is pressed for about 35 seconds.

- Data records can be locked to avoid inadvertent deletion.

- Two new AAA size batteries can be used for more than 500 times measurement; the battery power status is indicated 

on the screen.

- Small in size and light in weight to carry.

- Easy to operate, especially suitable for old people.

1.5 Scope of Application

The device is intended for self-monitoring and storing ECG waveform and heart rate.

The device is not the monitor used in clinical institution or hospital, but is used for spot-check purpose ONLY. 

It can not be used to substitute the normal ECG examination or real time monitoring. The measurement 

results are a useful reference for doctor, but do not make direct diagnostic or analytical decision based 

on the information provided by this device.

ENGLISH

Summary of Contents for Cardio-A Palm ECG

Page 1: ...ENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual befo...

Page 2: ...amination is suggested for normal examination of your heart condition 10 Connect Disconnect data cable to the monitor gently and carefully do NOT wring or shake the data cable plug with force to inser...

Page 3: ...prises main board LCD electrodes etc 1 4 Features Two input connections for ECG signal Embedded metal electrodes Lead wire connector optional Two output modes for ECG data LCD display Serial output in...

Page 4: ...ion the device will enter into a preparing phase this phase will last 6 seconds see the beeside picture during this period the device will judge whether the placed position is befitting Please adjust...

Page 5: ...ECG signal detected by Leg measurement is equivalent to Lead II ECG signal Improper operating method A You moved your foot B Wrong holding gesture C The electrode III can not contact bare skin comple...

Page 6: ...m is determined at the beginning of the measurement automatically 2 1 2 and 1 3 2 1 The measuring time lasts about 30 seconds Then enter the analysis phase in turn Note Refer to the period of result d...

Page 7: ...and press to measure ECG Refer to 3 2 for details 3 3 2 The function of waveform review 1 On the main menu screen press the key or of Navigation key to select Waveform Review and press to enter the re...

Page 8: ...esults and analysis Do not attempt self diagnosis 2 5 If the detected signal is too weak the beeside description can be seen Please try again 2 6 On the result description screen press the power back...

Page 9: ...he key or to move the cursor for selecting setting option and then press the key or to set numerical value 3 Press the power back button to complete the setting and return to the previous menu 3 3 5 2...

Page 10: ...ls Notes During uploading the keys and power key are effective After upload turn off the device and restart it again to start measurement Keep the monitor and data cable still during uploading data Un...

Page 11: ...tal electrode with medical alcohol when the measuring signal is too weak Please wipe the surface of the unit with clean cloth or let it dry in air Please take out the batteries if the unit will not be...

Page 12: ...etic interference 4 Weak signal Trouble ECG waveform excursion strong irrelevant waveform Solution 1 Clean skin with soap and water and eliminate skin crumb and grease to make the skin wet and greasel...

Page 13: ...rences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected products must be returned only to the d...

Page 14: ...dal Contraction Premature Ventricular Contraction or multi focus premature beat and help cure Testee s Symptom Normal heart beat is followed by a premature beat 4 Bigeminy It is a type of PVC in which...

Page 15: ...beat The display of missing beat may be doubted as Missed beats Symptom arrhythmia a long time interval Indication May be the conduction system of heart blocks it often occurs in functional cardiovas...