24

3.3.5.4 LCD Backlight Setting

1. On the system settings screen press the key 

 or 

 of Navigation key 

to choose Backlight Setting and press 

 to enter Backlight settings screen 

(see the following figure).

2. Press the key 

 or 

 of Navigation key to set backlight on/off.

3. Press the power/back button to complete saving and return to the previous 

menu.

4. On the others screen (except Waveform screen and Setting screen) short press the right key of Navigation key is 

able to turn on/off the backlight quickly.

3.3.5.5 Software Version

1. On the system settings screen press the key 

 or 

 of Navigation key to 

select Version and press 

 to enter Software Version screen.

2. Press the power/back button to return to the previous menu.

3.3.6 Upload ECG Data

1. On the system settings screen press the key 

 or 

 of Navigation key to 

select Upload ECG data and press  

  to enter into Upload ECG data screen.

On upload data screen, connect the monitor to PC with appropriative data 

cable, all the records of waveform recall list and waveform description list can 

be uploaded to PC for saving after entering into ECG Viewer System. Refer to 

ECG Viewer Manager User Manual for details.

Notes:

- During uploading, the keys 

  and power key are effective. After upload, turn off the device and restart it again to 

start measurement.

- Keep the monitor and data cable still during uploading data. Unplug the data cable plugs on the halfway of uploading 

data is not recommended.

- If the battery power is not enough, please change the batteries, or uploading operation will be interrupted.

This function is optional, and needs software support

(ECG Viewer Manager System).In addition, an appropriative data cable supplied by GIMA is required.

4. Technical Parameters

4.1 ECG Monitoring

- The number of channels: single (differential input).

- Electrodes: 3 embedded metal electrodes or using 3 adhesive ECG electrodes by connection to the lead wire.

- Body measurement: the detecting ECG signal can be the following lead selection:

Lead I (between right hand and left hand);

Chest Lead (between right hand and chest);

Lead II (between right hand and left leg).

- Signal bandwidth: 0.5Hz~40Hz

- Internal noise level: =30µVp-p

- Heart rate measuring range: 30bpm~240bpm

- Heart rate measuring precision: ±2bpm

- Display sensitivity: 5.0 mm/mV±10%

- Common-mode rejection ratio (CMRR): >=60dB

- Sweeping speed: 20mm/s±10%

- Input loop current: ≤0.1μA

- Input impedance: ≥MΩ

4.2 Display

- 128×64dot-matrix single color LCD

- Display dimension: 49mm×25mm

- Display of waveform: 128×48 pixels

- Backlight: LED backlight.

ENGLISH

Summary of Contents for Cardio-A Palm ECG

Page 1: ...ENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual befo...

Page 2: ...amination is suggested for normal examination of your heart condition 10 Connect Disconnect data cable to the monitor gently and carefully do NOT wring or shake the data cable plug with force to inser...

Page 3: ...prises main board LCD electrodes etc 1 4 Features Two input connections for ECG signal Embedded metal electrodes Lead wire connector optional Two output modes for ECG data LCD display Serial output in...

Page 4: ...ion the device will enter into a preparing phase this phase will last 6 seconds see the beeside picture during this period the device will judge whether the placed position is befitting Please adjust...

Page 5: ...ECG signal detected by Leg measurement is equivalent to Lead II ECG signal Improper operating method A You moved your foot B Wrong holding gesture C The electrode III can not contact bare skin comple...

Page 6: ...m is determined at the beginning of the measurement automatically 2 1 2 and 1 3 2 1 The measuring time lasts about 30 seconds Then enter the analysis phase in turn Note Refer to the period of result d...

Page 7: ...and press to measure ECG Refer to 3 2 for details 3 3 2 The function of waveform review 1 On the main menu screen press the key or of Navigation key to select Waveform Review and press to enter the re...

Page 8: ...esults and analysis Do not attempt self diagnosis 2 5 If the detected signal is too weak the beeside description can be seen Please try again 2 6 On the result description screen press the power back...

Page 9: ...he key or to move the cursor for selecting setting option and then press the key or to set numerical value 3 Press the power back button to complete the setting and return to the previous menu 3 3 5 2...

Page 10: ...ls Notes During uploading the keys and power key are effective After upload turn off the device and restart it again to start measurement Keep the monitor and data cable still during uploading data Un...

Page 11: ...tal electrode with medical alcohol when the measuring signal is too weak Please wipe the surface of the unit with clean cloth or let it dry in air Please take out the batteries if the unit will not be...

Page 12: ...etic interference 4 Weak signal Trouble ECG waveform excursion strong irrelevant waveform Solution 1 Clean skin with soap and water and eliminate skin crumb and grease to make the skin wet and greasel...

Page 13: ...rences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected products must be returned only to the d...

Page 14: ...dal Contraction Premature Ventricular Contraction or multi focus premature beat and help cure Testee s Symptom Normal heart beat is followed by a premature beat 4 Bigeminy It is a type of PVC in which...

Page 15: ...beat The display of missing beat may be doubted as Missed beats Symptom arrhythmia a long time interval Indication May be the conduction system of heart blocks it often occurs in functional cardiovas...

Reviews: