29
G. Environment requirement
Operating Temperature: 5 ~40°C
Operating Humidity: 30~80%
Atmospheric pressure: 70~106kPa
H. The performance under low perfusion condition
The accuracy of SpO
2
and PR measurement still meets the specifica-
tion described above when the pulse modulation amplitude is as low
as 0.6%.
I. Resistance to ambient light interference:
The accuracy of SpO
2
and PR measurement still meets the specifi-
cation described above when the device is tested by SpO
2
simulator
(Fluke Biomedical Index 2 series) while setting the emulating interfer-
ence of sun light and 50Hz/60Hz fluorescent light.
J. Dimensions:
60 mm (L) × 33 mm (W) × 30 mm (H)
Net Weight:
35g (including battery)
K. Classification:
The type of protection against electric shock: Internally powered
equipment.
The degree of protection against electric shock: Type BF applied
parts.
The degree of protection against harmful ingress of liquids: Ordinary
equipment without protection against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B.
6. ACCESSORIES
A. A lanyard
B. Two batteries
C. A carrying pouch
D. A User Manual
Note: The accessories are subject to change. Detailed items and
quantity see the Packing List.
ENGLISH