119
Immunity test
IEC 60601
test level1
Compliance
level
Electromagnetic
environment
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
N/A
N/A
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1: Guidance and manufacturer’s declaration - electromagnet-
ic emissions
The device is intended for use in the electromagnetic environment spec-
ified below. The user of this device should assure that it is used in such
an environment.
Emissions test
Compliance Electromagnetic environment
RF emissions
CISPR 11
Group 1
This device uses RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and are not likely to
cause any interference in nearby electron-
ic equipment.
RF emissions
CISPR 11
Class B
This device is suitable for use in all estab-
lishments. It uses internal power supply
and has no connection with power supply
network.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
Table 2: Guidance and manufacturer’s declaration - electromagnet-
ic immunity
The device is intended for use in the electromagnetic environment spec-
ified below. The user of this device should assure that it is used in such
an environment.