CAPNO CUBE CAPNOGRAPH

User manual

M33831

-GB-R

ev

.0-06.22

Imported by: Gima S.p.A. 

Via Marconi, 1 - 20060 Gessate (MI) Italy
[email protected] - [email protected]

www.gimaitaly.com

0482

IP33

Beijing Kingst Commercial & Trade Co., Ltd.
6th Floor, 2-7, Road Beibinhe in Xibianmen,
100045 Beijing, China
MADE IN CHINA

CAPNO-T 

(GIMA 33831)

SUNGO Europe B.V.
Olympisch Stadion 24, 1076DE Amsterdam, Netherlands

Summary of Contents for 33831

Page 1: ...A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com 0482 IP33 Beijing Kingst Commercial Trade Co Ltd 6th Floor 2 7 Road Beibinhe in Xibianmen 100045 Beijin...

Page 2: ...graph s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and storage etc as well as the safety proce...

Page 3: ...es on the product due to the technologies promotion or due to the special demands of user which do not influence the monitor s work will not be informed further Furthermore please pay attention to the...

Page 4: ...Monitor 13 Chapter 4 Patient connection 14 4 1 CO2 measurement 14 4 2 Respiration rate measurement 15 4 3 The sensor s zero 16 4 4 Notice 16 Chapter 5 Screen display and Operation 17 5 1 Screen main d...

Page 5: ...ble Shooting Analysis 26 Appendix 1 Explanations of Terms in this Manual 27 Appendix 2 ENGINEER MENU Changing compensation of balance gas 28 Appendix 3 Guidance and manufacturer s declaration Electrom...

Page 6: ...d due to misuse or incorrect operation i e without following the User manual instruction The monitor is damaged due to incorrect connection with another instrument The monitor is accidentally damaged...

Page 7: ...t from the monitor causing the monitor to fall on the patient Warning If the monitor falls off accidentally please do not operate it until its safety and performance have been carefully tested and pos...

Page 8: ...ient during MRI scanning Use during MRI scanning may cause a burn to the patient and affect the quality of MRI image or the measurement accuracy of the monitor Warning MR unsafe Do not expose the devi...

Page 9: ...mperes Warning Electrical installation of the room or the building in which the monitor is to be used must comply with regulations specified by the country in which the equipment is to be used Electro...

Page 10: ...o compensation should be set in the balance gas MENU to meet the accuracy requirements if such gases or vapours are present Note 2 The accuracy of CO2 concentration measurement is also influenced by R...

Page 11: ...Up Time 15 seconds Total system response time 1s Memory 24 hours on Screen Trend and Numeric Respiration Rate Range 3 150 rpm Accuracy 1 of reading or 1 rpm whichever is greater Memory 24 hours on Sc...

Page 12: ...x D Weight 80g including lithium battery and adult airway adapter Warranty Maintenance Calibration One year warranty on Main Unit and Lithium Ion Rechargeable Battery IP rating IP33 93 42 EEC Medical...

Page 13: ...b In the main screen if the bluetooth printer is equipped to press this button for two seconds to drive the printer to print CO2 reaction curve and result data of all parameters 4 Multifunction butto...

Page 14: ...attery compartment location 8 DC5V Mini USB Charging interface Note this interface must only be connected to a device which meets safety standards 9 Hanging Point for Lanyard if required 10 Airway ada...

Page 15: ...Modules can measure atmospheric pressure and atmospheric pressure can compensate the calculation for the concentrations of carbon dioxide which improve the design accuracy Then the monitor CO2 module...

Page 16: ...can make zero Then press button send zero order and ZEROING will be shown and wait for 15 20 seconds till ZEROING disappears 4 4 Notice Caution Under the conditions of electromagnetic influence for ex...

Page 17: ...OTO mode please see the details in 5 5 NEW PATIENT b The symbol appears if the bluetooth module is equipped If this symbol is green it indicates that no bluetooth equipment is connected e g bluetooth...

Page 18: ...use potential danger in isolated area like ICU or operating room Please refer to the content of menu setting of CO2 It is critical to set alarm of physiological parameter which gives alarm clinical si...

Page 19: ...t icon becomes two minutes later trumpet icon becomes meanwhile alarm begins to work if there is sound alarm If to press the button in this period the silence alarm can be released Silence alarm time...

Page 20: ...utton to enter the next submenu to press again to go back to the main display screen This menu includes the following options The setting menu for CO2 CO2_SETUP The time menu TIME_SETUP The new patien...

Page 21: ...EA TIME 15s 44s off 6 The unit of CO2 CO2 UNIT mmHg or kPA 7 Sensor zero Calibration 8 CO2 Wave scale WAVE SCALE 54mmHG or 76mmHG 9 Default reload LOAD DEFAULTS 10 Exit EXIT Attention a When respirati...

Page 22: ...justment The procedure is as follows 1 Change time 2 Move the cursor to SAVE then press the button to enter the following menu FIGURE 8 3 YES is already selected highlighted in white and if you wish t...

Page 23: ...the original data of the same ID of patient 5 EXIT to quit the current menu but not to store the changed setup Chapter 6 Charging Maintenance Cleaning 6 1 Charge Connect the AC DC power adapter via th...

Page 24: ...usage every time and check in fared window surface clean condition 6 3 Cleaning Warning Before to clean the monitor power shall be turned off and removed from any charging source 1 Cleaning the Monit...

Page 25: ...can cause cross infection For reusable airway adapter rA3 Soak the adapter in disinfectant and dry it in a ventilated place for at least 30min after taking it out See the following table for specific...

Page 26: ...oncentration is zero Show NO ADAPTER or SENSOR ERR or IR LAMPBAD onthescreen 1 Without adapter 2 Sensor data wrong 3 Light source wrong 1 Check adapter if plugged in 2 Check adapter if plugged into co...

Page 27: ...abnormally 1 To check the AC DC adapter and cable Attention Please to contact the client service centre if some problems occurred repeatedly Appendix 1 Explanations of Terms in this Manual MENU Menu...

Page 28: ...ensation CALIBRATE Calibration CANCEL Cancellation Appendix 2 ENGINEER MENU Changing compensation of balance gas Attention Only the trained personnel may carry out the following the procedure Contact...

Page 29: ...is menu can be directly adjusted such as LOAD DEFAULT or EXIT to press button to execute This menu including the following setups BARO PRESS 760mmHg BALANCE GAS AIR N2O and HELIUM O2 CONCENTRATION 20...

Page 30: ...n such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 This device uses RF energy only for its internal function Therefore its RF emissions...

Page 31: ...15kV air Floors should be wood concrete or ceramic tile if floors are covered with synthetic material the relative humidity should be at least 30 If ESD interfere with the operation of equipment count...

Page 32: ...it is recommended that the equipment or system be powered from an uninterruptible power supply or a battery Power frequency 50Hz 60Hz magnetic field IEC61000 4 8 30A m 30 A m Power frequency magnetic...

Page 33: ...of this device including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance 1 2 P 1 2 P 80MHz...

Page 34: ...to the maximum output power of the communications equipment Rated maximum output power of transmitter W Distanza di separazione in base alla frequenza del trasmettitore m 150 kHz to 80 MHz 80 MHz to...

Page 35: ...ric Airway Adapter Optional 1 Neonatal Pediatric Airway Adapter Optional 1 USB Cable 1 AC DC Adapter 1 Symbol Date of manufacture Medical Device complies with Directive 93 42 EEC Serial number Expirat...

Page 36: ......