20

PREPARATION

1. Pull out the air hose from the nebulizer leaving it plugged to the air outlet pipe of the 

device.

2.  Rotate the upper part of the nebulizer anticlockwise.

3.  Disconnect the internal pisper at the bottom of the nebulizer with finger force.

CLEANING

Clean all the components of the nebulizer (except the air hose) before and after each use by 

choosing one of the methods described below.

Method 1: 

Thoroughly clean the components for 5 minutes by using hot drinking tap water 

(around 40°C – 104°F) and neutral soap.

Method 2: 

Clean the components (except the air hose) by immersion in a solution of 60% 

of water and 40% of white vinegar. At the end of the operation, rinse with plenty of hot 

drinking water (around 40°C – 104°F).

At the end of the cleaning operations, rinse thoroughly by removing excess water and allow 

to air-dry in a clean spot.

DO NOT BOIL OR AUTOCLAVE THE AIR HOSE AND THE MASKS

DO NOT WASH THE ACCESSORIES IN THE DISH WASHER

WASHING

Where diseases with risk of infection and microbial contamination are present, the end user 

shall carry out the washing operations properly. The washing procedure can be performed 

only if the components have been previously cleaned (see cleaning section).

For the washing procedure, the following operations shall be carried out:

-  Fill a container of suitable size to hold every individual component with drinking water and 

disinfectant (hypochlorite solution easily available in pharmacies) by complying with the 

proportions indicated on the disinfectant packaging.

-  The immersion time in the solution is indicated on the packaging of the hypochlorite solu-

tion according to the concentration chosen for the preparation of the solution.

- Rinse with plenty of warm drinking water until removing any trace of solution, dry and 

store in a dry and dust-free place.

-  Dispose the solution used according to the instructions provided by the manufacturer of 

the disinfectant. 

SUPPLIED ACCESSORIES

ACCESSORIES 

HI-FLO accessories kit 

(HI-FLO ampoule, Adult Mask, Children Mask, 

Air hose, Mouthpiece and Nosepiece)

Air filter (+2 spare parts)

Transport bag

ENGLISH

Summary of Contents for 28102

Page 1: ...the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ACHTUNG Die Bediener m ssen vorher dieses Handbuch gelesen und verstanden haben bevor sie das Pro...

Page 2: ...nected 3 Do not leave the appliance plugged unnecessarily disconnect the plug from the power network if not used 4 Comply with the safety standards for electrical appliances In particular Use only ori...

Page 3: ...people with disabilities 10 Do not leave the device unattended in areas accessible to minors and or people with limited mental capacity as they may be strangled with the air hose 11 The patient can c...

Page 4: ...to any of its part due to accident misuse and or abuse Any unauthorized intervention on the device even the slightest one will imme diately invalidate the warranty and does not ensure the compliance w...

Page 5: ...detergents Do not use excessively wet cloths as the contact of liquids with the electri cal parts of the appliance may cause malfunctions or may be hazardous While cleaning the device make sure that n...

Page 6: ...ntact the reseller or Gima s technical service BEFORE CARRYING OUT ANY CHECKS IN THE EVENT OF ANOMALIES OR MALFUNCTIONS CONTACT GIMA S TECHNICAL SERVICE THE MANUFACTURER OFFERS NO GUARANTEE FOR THE AP...

Page 7: ...WASHING Where diseases with risk of infection and microbial contamination are present the end user shall carry out the washing operations properly The washing procedure can be performed only if the co...

Page 8: ...or if the device is no longer efficient If the filter is too dirty it shall be replaced AIR FILTER REPLACEMENT The air filter shall be replaced every 25 hours of operation or when it is too dirty For...

Page 9: ...e nebulizer facing upwards in order to prevent any substances and or the medicine from leaking out of the nebulizer during normal use In the event of overfilling empty the ampoule clean it and repeat...

Page 10: ...e electromagnetic environment Radiated emissions Conductions CISPR11 Group 1 The Mistral aerosol uses RF energy only for Internal function Therefore its RF emissions are very low and do not cause any...

Page 11: ...ower of the transmitter in Watt W according to the manufacturer of the transmitter and d is the recommended separation distance calculated in meters m The field strengths from fixed RF trans mitters a...

Page 12: ...eparation distances between portable and mobile radio communi cation appliances and the monitor The Mistral aerosol is designed to operate in an electromagnetic environment in which the RF radiated in...

Page 13: ...expenses and packaging not included All components subject to wear are not included in the warranty The repair or replacement performed during the warranty period shall not extend the warranty The wa...

Reviews: