24
ENGLISH
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-5
Surge
IEC 61000-4-5
Voltage dips,short
interruptions and
voltage variations
on power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 HZ) magnetic
field
IEC 61000-4-8
IEC 60601
Test level
±2 KV contact
±4 KV contact
±6 KV contact
±2 KV air
±4 KV air
±8 KV air
±2 KV for power
supply lines
±1 KV
differential mode
±2 KV common
mode
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60% dip in U
T
)
for cycle
70% U
T
(30% dip in U
T
)
for cycle
<5% U
T
(>95% dip in U
T
)
for 5 s
3A/m
Electromagnetic environment
guidance
Floor should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at less 30 %
Mains power quality should be
that of a typical commercial
or hospital environment
Mains power quality should be
that of a typical commercial
or hospital environment
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of 27354
requires continued operation during
power mains interruptions, it is
reccomended that the 27354 be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic field
should beat levels characteristic
of a typical location in a typical
commercial or hospital environment.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The equipment
27354
is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment
27354
should assure that it is used in such an environment
Compliance
level
±2 KV contact
±4 KV contact
±6 KV contact
±2 KV air
±4 KV air
±8 KV air
±2 KV for power
supply lines
±1 KV
differential mode
±2 KV common
mode
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60%dip in U
T
)
for cycle
70% U
T
(30% dip in U
T
)
for cycle
<5% U
T
(>95% dip
in U
T
) for 5 s
3A/m