STERLIZATION PROCEDURE.
The twin insuffl ator can treated by surface disinfection.
2. Disinfection:
Soaking in solutions or Thermo chemically in a washer Sterilizer up to 93°C maximum may
perform disinfection. Manufacturer’s instruction regarding duration and concentration of
solutions should be strictly adhered. After Disinfection, rinse thoroughly in sterile water and
dry with a clean lint free cloth.
3. Cold Soak Solution:
To achieve a high-level disinfection, Cidex®OPA or 2.4% Glutaraldehyde solution may be used
according to manufacturer’s instructions. Dry with lint free, clean cloth or fi ltered pressurized
air. Do not immerse F.O. Instrument in Bleach, Betadine or Potassium Hydroxide solutions.
Doing so will several damage instruments also avoid metal to metal contact after soaking, the
F.O. Instrument should be rinsed under sterile water to remove chemical residues and dry with
lint free clean cloth or fi ltered pressurized air.
4. Sterilization:
Before performing any of the procedures described below, F.O. Instrument Head should be
cleaned as described in the cleaning Procedure.
5. Gas Sterilization:
Gas sterilization by Ethylene oxide up to a maximum temperature of 65°C and 8 psi may be
performed, which is preferred especially if sterilization is to be performed regularly.
6. Autoclave:
In order to perform Autoclave kindly refer to below mentioned table:
Note:
Do not exceed temperature of 135°C and pressure of 28 psi Flash autoclaving and hot air
sterilization should be avoided as these processes will damage the instrument.
9
ENGLISH
(A) GRAVITY
DISPLACEMENT STEAM
(B) PRE-VACUUM STEAM
Temperature
Cycle Time
Dry Time
121 °C (250 F)
30 Minutes
15 Minutes
134 °C (270 F)
5 Minutes
20 Minutes