APPENDIX RESET PROCEDURE
When “Init 1”, “Init 2” or “Init 3” appears on display
please follow this procedure to reset the pipette to
the factory calibration settings.
1) Unlock
or
Lock/Unlock
(if the init message
a p p e a r s i n u n l o c k
position) the volume
(push the thumbwheel
upwards), the pipette
displays “turn –”.
2) Rotate the thumbwheel clockwise to decrease
the volume to the mechanical stop ( the thumb-
wheel will continue to turn but the display will
not change).
3) Lock the volume (push
the thumbwheel down-
wards), the pipette dis-
plays “turn +”.
4) Unlock
the
volume.
5) Rotate the thumbwheel
counterclockwise to
increase the volume
until “good” appears on
display.
6) Lock the volume.
7) The pipette is ready to operate (operating volume
on display).
After resetting, the pipette is correctly adjusted to the factory
calibration settings.