30
PIPETMAN
®
u
ltra
For Single-channel
Take 10 measurements using distilled water at the
specifi ed readjustment volume.
For Multichannel
Take 10 measurements from the fi rst channel (1) and
10 from the last channel (8 or 12). Calculate the mean
volume from the 20 measurements.
For example, for an Ultra 8x300 the recalibration
volume is 60 μL. You set the LCD to show 60 μL, but
you calculate a mean volume of 58 μL; a mean error
of -2 μL. Reset the pipette to show:
Recalibration Volume
minus
Mean Error,
in this example 60 - (-2) = 62 μL.
Open the battery compartment, and plug in the
user-calibration tool (as shown) to reset the displayed
volume to 60 μL.
The LCD will show RECAL in the volume setting window. This
indicates that the factory calibration has been modifi ed. To
remove the RECAL sign and obtain a “factory” calibration, send
the pipette back to your Gilson distributor.
Remove the user-calibration
tool, close the battery com-
partment, and calculate the
accuracy as described in
“Procedure for Evaluating
Accuracy and Precision of
Gilson Pipettes”, which is
Readjustment Balance
Diamond
Model
Volume (μL)
Sensitivity (g)
Tips
U2 0.5
10
-6
DL10
U10 1
10
-6
DL10
U20 2
10
-6
D200
U100 20
10
-5
D200
U200 50
10
-5
D200
U1000 200
10
-4
D1000
U5000 1000
10
-4
D5000
U10 ml
1000
10
-4
D10ml
8x20 and 12x20
2
10
-6
DL10
8x300 and 12x300
60
10
-5
D300
User-calibration tool