background image

5

4

2 - PARTS CHECK LIST

Just take a moment to verify that the following items 

are present:

  MICROMAN E,

   User's Guide,

   Safety bag,

   Adhesive id-tags (strip of 6),

   Capillary pistons (10),

   Certificate of conformity

 

(including barcode sticker).

3 - DESCRIPTION

1

  Color-coded push-button and 

magnifier with volume range 

for aspirating and dispensing.

2

 

Body or handle.

3

 Capillary.

 Piston.

  Shaft.

  Thumbwheel for 

setting the vol-
ume.

 Volumeter.

 

Volume unlock button.

 

Data matrix.

Never lubricate the capillary holder, capillary, 
or any other part of the pipette. If the capillary 
tends to slip off the capillary holder, clean the 
capillary holder with ethanol, using a medical 
wipe or similar soft tissue.

CP 

Capillary Piston

  

(C) 

(P)

CP10

 

Polypropylene Polyester

CP25

 

Polypropylene Polyester

CP50

 

Polypropylene Polyester

CP100

 

Polypropylene Polyethylene

CP250

 

Polypropylene Polyethylene

CP1000

 

Polypropylene Polyacetal

5 - 

FITTING THE CAPILLARY AND PISTON

4 - OPERATING RANGES AND MATERIAL

Available Models

Model 

Volume Range

M10E 

 

1 µL to 10 µL

M25E

 

3 µL to 25 µL

M50E

 

20 µL to 50 µL

M100E

  

10 µL to 100 µL

M250E

 

50 µL to 250 µL

M1000E

 

100 µL to 1000 µL

Materials

 

Model Spring Clamp  Shaft

 Body 

  

 

 

 

(B) 

M10E

 

SS 

Be Alloy 

PBT 

PVDF/PP 

M25E

 

SS 

Be Alloy 

PVDF 

PVDF/PP 

M50E

 

SS 

Be Alloy 

PVDF 

PVDF/PP 

M100E

 

SS 

Be Alloy 

PBT 

PVDF/PP 

M250E

 

SS 

Be Alloy 

PVDF 

PVDF/PP 

M1000E

 

SS 

Be Alloy 

PVDF 

PVDF/PP 

 

  Abbreviations

SS = Stainless Steel
Be = Beryllium
PBT = Polybutylene Terephthalate
PP = Polypropylene
PVDF = Polyvinylidene Fluoride

8

Summary of Contents for MICROMAN E M1000E

Page 1: ...Ergonomic Pipetting for Problem Liquids ...

Page 2: ...hese are the only parts to be in touch with the aspiratedliquids andreplacementthereofwillprovide absolute protection against contamination caused by carryover between samples or reagents Simply replace the capillary and piston between assays they are ejected automatically and simultaneously thus avoiding any risk for the operator MICROMAN E is equipped with a direct reading volumeter that allows ...

Page 3: ...ing a medical wipe or similar soft tissue CP Capillary Piston C P CP10 Polypropylene Polyester CP25 Polypropylene Polyester CP50 Polypropylene Polyester CP100 Polypropylene Polyethylene CP250 Polypropylene Polyethylene CP1000 Polypropylene Polyacetal 5 FITTING THE CAPILLARY AND PISTON 4 OPERATING RANGES AND MATERIAL Available Models Model Volume Range M10E 1 µL to 10 µL M25E 3 µL to 25 µL M50E 20 ...

Page 4: ...lack or red to indicate the position of the decimal point according to the model see examples 3 2 MICROMAN E CP model CP10 M10E CP25 M25E CP50 M50E CP100 M100E CP250 M250E CP1000 M1000E CP10 CP25 CP50 CP100 CP250andCP1000Tipack Capillary pistons CP are deliveredreadytouse Tofit a capillary piston proceed as follows 1 Gently press the MICROMAN E onto the capillary piston until it is firmly seated o...

Page 5: ... the capillary piston Press the push button to the first stop 1 then press down harder to the second stop 2 at this point the piston and capillary are ejected simultaneously MICROMAN E is automatically calibrated by correctly fitting a capillarypiston Once thevolumeisset measurementswillbe accurate and precise 8 GLP FEATURES No need to touch disposables capillary piston Serial Number engraved on p...

Page 6: ...haft 3 Set aside the body handle in a dry and secure location 4 Clean the individual components using an ultrasonic bath for 20 minutes at 50 C or with a soft cloth and brush Small round brushes with soft plastic bristles may be used to clean the interior of the shaft 5 Rinse the individual components with distilled water 6 Leave the parts to dry by evaporation or wipe them down with a clean soft ...

Page 7: ...sis of adjustment Ex Reference temperature 20 C Relative humidity 50 Barometric pressure 101 kPa Use of distilled water grade 3 ISO 3696 Ten measurements for each test volume which are nominal volume 50 of nominal volume and the minimum or 10 of nominal volume Performance Tests Eachpipetteisinspectedandvalidatedaccordingtothe Gilson Quality Assurance System Based on extensive historicaldata manufa...

Page 8: ...e purchase price paid for such pipettes THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMSALLOTHERWARRANTIES WHETHEREXPRESSORIMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCESSHALLGILSONBELIABLEFORANYCONSEQUENTIAL PUNITIVE INDIRECTORINCIDENTALDAMAGESARISINGOUTOFANY BREACH OF ANY EXPRESS OR IMPLIEDWARRANTY E...

Page 9: ...ilson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 I Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I BP 145 F 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 I Fax 33 0 1 34 29 50 20 www gilson com sales gilson com service gilson com training gilson com ...

Reviews: