9

8

- when 

increasing 

the volume setting, pass the required 

value by 1/3 of a turn and then slowly decrease the 

volume to reach the required setting, making sure not 

to overshoot the mark.

For large volume changes, 

proceed 

as follows :
1) Push the unlock button counter.
2) When 

decreasing 

the volume 

setting, approach the required 

setting, by 1/3 of a turn.

 When 

increasing 

the volume 

setting, pass the required setting 

by 1/3 of a turn.

3) Release the unlock button counter.

7 - PIPETTING

Aspirating

-  Press the push-button to the 

first stop (1),

-  Immerse the capillary 2 mm 

into the liquid,

Slowly release the push-

button to draw up the liquid 

(top position),

-  Wipe any liquid from the out-

side of the capillary, taking care 

not to touch the orifice.

Dispensing

-  Place the end of the capillary against the inside wall 

of the recipient vessel,

-   Press the push-button slowly to the first stop,
-  Keeping the push-button depressed, move the capil-

lary away from the sidewall,

- Withdraw MICROMAN E from the vessel and release 

the push-button.

Ejecting the capillary piston

Press the push-button to the first stop (1), then press 

down harder to the second stop (2); at this point the 

piston and capillary are ejected simultaneously.

MICROMAN E is automatically calibrated by correctly fitting a 
capillary piston. Once  the volume is set, measurements will be 
accurate and precise.

8 - GLP FEATURES

-  No need to touch disposables (capillary piston).
-  Serial Number: engraved on pipette body.
-  Barcode: on the box and with the certificate (can be 

transferred).

-  ID Tag (Application or User).
-  Color-coded push-button.
-  Color-coded magnifier with volume range.
-  Data matrix engraved on pipette body.
-  Gilson certificate of conformity according 

to ISO 8655.

9 - TROUBLESHOOTING

You may be able to identify and correct the problem 

by referring to the following table. If you cannot solve 

the problem, please contact your Gilson representative.

Leaks:

 

Replace the capillary and piston.

No stroke:

 

The piston is not properly fitted into the clamp; check 

that the piston is seated correctly (refer to pages 5-7). 

Without any capillary piston, press the push-button to 

the first stop, then press down harder to the second stop.

Difficult to fit a capillary:

 

Clean the capillary holder with ethanol. Without any 

Top

1
2

Summary of Contents for MICROMAN E M1000E

Page 1: ...Ergonomic Pipetting for Problem Liquids ...

Page 2: ...hese are the only parts to be in touch with the aspiratedliquids andreplacementthereofwillprovide absolute protection against contamination caused by carryover between samples or reagents Simply replace the capillary and piston between assays they are ejected automatically and simultaneously thus avoiding any risk for the operator MICROMAN E is equipped with a direct reading volumeter that allows ...

Page 3: ...ing a medical wipe or similar soft tissue CP Capillary Piston C P CP10 Polypropylene Polyester CP25 Polypropylene Polyester CP50 Polypropylene Polyester CP100 Polypropylene Polyethylene CP250 Polypropylene Polyethylene CP1000 Polypropylene Polyacetal 5 FITTING THE CAPILLARY AND PISTON 4 OPERATING RANGES AND MATERIAL Available Models Model Volume Range M10E 1 µL to 10 µL M25E 3 µL to 25 µL M50E 20 ...

Page 4: ...lack or red to indicate the position of the decimal point according to the model see examples 3 2 MICROMAN E CP model CP10 M10E CP25 M25E CP50 M50E CP100 M100E CP250 M250E CP1000 M1000E CP10 CP25 CP50 CP100 CP250andCP1000Tipack Capillary pistons CP are deliveredreadytouse Tofit a capillary piston proceed as follows 1 Gently press the MICROMAN E onto the capillary piston until it is firmly seated o...

Page 5: ... the capillary piston Press the push button to the first stop 1 then press down harder to the second stop 2 at this point the piston and capillary are ejected simultaneously MICROMAN E is automatically calibrated by correctly fitting a capillarypiston Once thevolumeisset measurementswillbe accurate and precise 8 GLP FEATURES No need to touch disposables capillary piston Serial Number engraved on p...

Page 6: ...haft 3 Set aside the body handle in a dry and secure location 4 Clean the individual components using an ultrasonic bath for 20 minutes at 50 C or with a soft cloth and brush Small round brushes with soft plastic bristles may be used to clean the interior of the shaft 5 Rinse the individual components with distilled water 6 Leave the parts to dry by evaporation or wipe them down with a clean soft ...

Page 7: ...sis of adjustment Ex Reference temperature 20 C Relative humidity 50 Barometric pressure 101 kPa Use of distilled water grade 3 ISO 3696 Ten measurements for each test volume which are nominal volume 50 of nominal volume and the minimum or 10 of nominal volume Performance Tests Eachpipetteisinspectedandvalidatedaccordingtothe Gilson Quality Assurance System Based on extensive historicaldata manufa...

Page 8: ...e purchase price paid for such pipettes THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMSALLOTHERWARRANTIES WHETHEREXPRESSORIMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCESSHALLGILSONBELIABLEFORANYCONSEQUENTIAL PUNITIVE INDIRECTORINCIDENTALDAMAGESARISINGOUTOFANY BREACH OF ANY EXPRESS OR IMPLIEDWARRANTY E...

Page 9: ...ilson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 I Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I BP 145 F 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 I Fax 33 0 1 34 29 50 20 www gilson com sales gilson com service gilson com training gilson com ...

Reviews: