7

6

Make sure the capillary piston (CP) 

model matches the MICROMAN E 

you are about to use: 

CP25, CP50 and CP250 Bulk

The capillaries and pistons are deliv-

ered in two separate boxes. To fit a 

capillary piston, proceed as illustrated:
1) Take care not to damage the 

sealing tip 

 when handling 

the piston 

.

2) Select a piston 

 and slide 

it into the capillary 

.

3) Gently push the capillary 

until it snaps onto the capil-

lary holder.

4) Slowly press the push-

button to the first stop to 

attach the stem 

. You 

should definitely feel and 

hear a slight click.

button to the first stop until you feel and hear a slight 
click 

3

4)  Do not release the push-button and pipette your 

solution.

6 - SETTING THE VOLUME

The volume of liquid to be aspirated is set using the 

volumeter. The volumeter consists of three number dials, 

which are read from top (most significant digit) to bottom 

(least significant digit). A marker is used to set exact or 

intermediate volumes using the scale on the bottom dial.

The dials are colored either black or red to indicate the 

position of the decimal point, according to the model (see 
examples).

3

2

 

MICROMAN E

CP model

CP10 M10E
CP25 M25E
CP50 M50E
CP100 M100E
CP250 M250E
CP1000 M1000E

CP10, CP25, CP50, CP100,

CP250 and CP1000 Tipack

Capillary pistons (CP) are 

delivered ready to use. To fit 

a capillary piston, proceed 

as follows:
1) G e n t l y   p r e s s   t h e 

MICROMAN E onto the 
capillary piston until it 
is firmly seated on the 
MICROMAN E 

.

2)  Pick up the capillary from 

the rack 

2

.

3) Slowly press the push-

Model

 

Color of volumeter numbers

 

Black Red 

Increment

M10E, M25E, M50E 

µL 

0.1 µL 

0.02 µL

M100E, M250E 

µL 

none 

0.2 µL

M1000E 

0.1 and 0.01 mL 

mL 

2 µL

3

2

The volume is set by turning the thumb-

wheel slowly to reach the required setting.

For small volume changes, 

directly and 

slowly turn the thumbwhell.
To obtain maximum accuracy when 

setting the volume, proceed as follows:
- when 

decreasing 

the volume setting, 

slowly reach the required setting, mak-

ing sure not to overshoot the mark.

8

3

6

8

36.8 µL

M50

0

6

6.8 µL

M25

1

6

8

168 µL

M250

0

7

5

0.75 mL

M1000

E

E

E

E

0

6

8

6.8 µL

M10

0

6

8

68 µL

M100

E

E

0

6

8

6.8 µL

M10

0

6

8

68 µL

M100

E

E

8

3

6

8

36.8 µL

M50

0

6

6.8 µL

M25

1

6

8

168 µL

M250

0

7

5

0.75 mL

M1000

E

E

E

E

Summary of Contents for MICROMAN E M1000E

Page 1: ...Ergonomic Pipetting for Problem Liquids ...

Page 2: ...hese are the only parts to be in touch with the aspiratedliquids andreplacementthereofwillprovide absolute protection against contamination caused by carryover between samples or reagents Simply replace the capillary and piston between assays they are ejected automatically and simultaneously thus avoiding any risk for the operator MICROMAN E is equipped with a direct reading volumeter that allows ...

Page 3: ...ing a medical wipe or similar soft tissue CP Capillary Piston C P CP10 Polypropylene Polyester CP25 Polypropylene Polyester CP50 Polypropylene Polyester CP100 Polypropylene Polyethylene CP250 Polypropylene Polyethylene CP1000 Polypropylene Polyacetal 5 FITTING THE CAPILLARY AND PISTON 4 OPERATING RANGES AND MATERIAL Available Models Model Volume Range M10E 1 µL to 10 µL M25E 3 µL to 25 µL M50E 20 ...

Page 4: ...lack or red to indicate the position of the decimal point according to the model see examples 3 2 MICROMAN E CP model CP10 M10E CP25 M25E CP50 M50E CP100 M100E CP250 M250E CP1000 M1000E CP10 CP25 CP50 CP100 CP250andCP1000Tipack Capillary pistons CP are deliveredreadytouse Tofit a capillary piston proceed as follows 1 Gently press the MICROMAN E onto the capillary piston until it is firmly seated o...

Page 5: ... the capillary piston Press the push button to the first stop 1 then press down harder to the second stop 2 at this point the piston and capillary are ejected simultaneously MICROMAN E is automatically calibrated by correctly fitting a capillarypiston Once thevolumeisset measurementswillbe accurate and precise 8 GLP FEATURES No need to touch disposables capillary piston Serial Number engraved on p...

Page 6: ...haft 3 Set aside the body handle in a dry and secure location 4 Clean the individual components using an ultrasonic bath for 20 minutes at 50 C or with a soft cloth and brush Small round brushes with soft plastic bristles may be used to clean the interior of the shaft 5 Rinse the individual components with distilled water 6 Leave the parts to dry by evaporation or wipe them down with a clean soft ...

Page 7: ...sis of adjustment Ex Reference temperature 20 C Relative humidity 50 Barometric pressure 101 kPa Use of distilled water grade 3 ISO 3696 Ten measurements for each test volume which are nominal volume 50 of nominal volume and the minimum or 10 of nominal volume Performance Tests Eachpipetteisinspectedandvalidatedaccordingtothe Gilson Quality Assurance System Based on extensive historicaldata manufa...

Page 8: ...e purchase price paid for such pipettes THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMSALLOTHERWARRANTIES WHETHEREXPRESSORIMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCESSHALLGILSONBELIABLEFORANYCONSEQUENTIAL PUNITIVE INDIRECTORINCIDENTALDAMAGESARISINGOUTOFANY BREACH OF ANY EXPRESS OR IMPLIEDWARRANTY E...

Page 9: ...ilson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 I Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I BP 145 F 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 I Fax 33 0 1 34 29 50 20 www gilson com sales gilson com service gilson com training gilson com ...

Reviews: