MetSet Editing Pages.
Feature and parameter available to view and select will depend on the MaxiMet model
connected. The following example pages refer to the GMX600 device. See page 5 for a
summary of available sensor outputs and parameters.
NOTE: Items shown in Bold are the default settings.
The Comms Page
can be used to Select:-
Protocol:-
GILL
, MODBUS or SDI-12 (for SDI-12 also set MetSet
Comms/Interface to SDI-12).
Interface:-
RS232
, RS422, RS485P2W (point to point) ,SDI-12 and
EXT.
EXT setting means RS422 or RS232 selected by a wire link
on connector pin 8 (see Page 19 table).
Baud Rate:-
4800, 9600,
19200
, 38400 or 57600
(1200 auto selected with SDI-12 setting).
Node ID:-
A to P,
Q
to Z.
Output Rate:-
0.25Hz to
1Hz
.
Message Mode:-
CONT
(Continuous output
)
or POLL (Polled Mode).
NOTE: When COMMS or Baud Rate settings are changed and Saved to Device a warning is
issued by MetSet. These setting changes will not become active until the MaxiMet power is
turned off and turned on again. The connecting device will then also need its Comms and
Baud rate settings changed to match the MaxiMet
.
The ASCII Set up Page
can be used to select:-
Termination:-
CRLF
or CR
Echo:-
ON
or OFF.