HS-50 & HS-100
Doc. No 1199-PS-0035 Issue 3
June 2021
___________________________________________________________________________________________
Page 6
6 TECHNICAL SPECIFICATION
* Accuracy specification applies for wind speeds less than 32m/s and for wind incidence < ±150° in the
horizontal plane and up to ±50° from the horizontal.
Parameter
HS-50
HS-100
Outputs
Data Output rates
From 0.4 to 50Hz
From 0.4 to 100Hz
Measurement Rate
50Hz
(Full 3 axis measurement)
100Hz
(Full 3 axis measurement)
Format
UVW or Polar (ASCII or Binary)
UVW or Polar (ASCII or Binary)
Wind Speed
Range
0 - 45 m/s
0 - 45 m/s
Resolution
0.01m/s
0.01m/s
Wind Speed Accuracy *
<1%RMS @ 12m/s
<1%RMS @ 12m/s
Wind Speed Offset
<±0.01ms
<±0.01ms
Direction
Range
0 – 359
0 – 359
Resolution
1
0
1
0
Directional Accuracy *
<±1
RMS
<±1
RMS
Speed of Sound Measurement
Measurement Rate
50Hz
100Hz
Range
300 -370 m/s
300 -370 m/s
Resolution
0.01 m/s
0.01 m/s
Accuracy *
<
0.5% (For wind speeds <30ms)
<
0.5% (For wind speeds <30ms)
Supply Input
Supply Voltage
9 to 30VDC
9 to 30VDC
Anemometer Supply Current
<300mA for 12VDC Input
<150mA for 24VDC Input
<300mA for 12VDC Input
<150mA for 24VDC Input
Environmental
Operating
Temperature
-40
°
C to +60
°
C
-40
°
C to +60
°
C
Storage Temperature
-50
°
C to +75
°
C
-50
°
C to +75
°
C
Altitude
0 to 3000m
0 to 3000m
Moisture/Dust Protection
IP65
IP65
Humidity
5 % to 100 %
5 % to 100 %
Precipitation
Operation up to 300 mm / hour
Operation up to 300 mm / hour
EMC
Emissions BS EN 50081 -1:1992
Immunity BS EN 50082 -1:1992
Emissions BS EN 50081 -1:1992
Immunity BS EN 50082 -1:1992
Input/Output Protection
>0.8J
>0.8J
Serial Interface
Electrical Interface
Full Duplex RS422
Full Duplex RS422
Baud Rates
2400, 4800, 9600, 19200, 38400,
57600, 115200
2400, 4800, 9600, 19200, 38400,
57600, 115200
Format
8 data bits, 1 stop bit, No Parity
8 data bits, 1 stop bit, No Parity
Analogue Inputs
Sampling Rate: measurement
50Hz Synchronous to wind
100Hz Synchronous to wind
Diff. Input Voltage Range
-5.000V to +5.000V
-5.000V to +5.000V
Accuracy
>0.05% of full scale
>0.05% of full scale
Resolution
14 bits
14 bits
Common Mode Input Range
>±8v
>±8v
Input Impedance
>1MΩ (Differential)
>1MΩ (Common Mode)
>1MΩ (Differential)
>1MΩ (Common Mode)
PRT
Resolution
0.01
°
C
0.01
C
Measurement Accuracy
<±0.1
°
C (-40 to +50
°
C)
<±0.1
°
C (-40 to +50
°
C)
<±0.15
°
C (+50 to +60
°
C)
<±0.15
°
C (+50 to +60
°
C)
Anemometer Sync Output
RS422 synchronous to Wind Measurement
RS422 synchronous to Wind Measurement
Inclinometer
Range
Monotonic over ±30
°
of tilt
Monotonic over ±30
°
of tilt
Resolution
0.01
°
0.01
°
Accuracy
±0.3
°
for tilt angles between ±10
°
±0.3
°
for tilt angles between ±10
°
Null Repeatability
±0.15
°
±0.15
°
Temperature Coefficient-Null
0.02
°
per
°
C
0.02
°
per
°
C
Temperature Coefficient –Scale Factor
0.2% per
°
C
0.2% per
°
C
Summary of Contents for 1199-PK-085