7.11 Cabling
Ensure that strain relief measures are employed when installing the cables. Do not allow the whole
weight of a cable to be applied to the connector.
An RS422 compatible cable should be used, with the number of twisted pairs matching the
application.
Generic description
– Twisted pairs with drain wire, screened with aluminised tape, with an overall
PVC sheath.
Typical Wire size 7/0.2mm (24AWG), 22AWG wire may also be used.
The table shows some suitable manufacturers’ references; other manufacturers’ equivalents can be
used.
No. of pairs
Gill ref.
Belden ref.
Batt electronics ref.
9
026 - 02663
8774
91009
The cable allows for the connection of all: -
Power inputs.
Digital data inputs and outputs.
Other example twisted pair screened cable types (24AWG) are:-
No. of pairs
Gill ref.
Belden ref.
Batt electronics ref.
2
-
9729
-
3
026 - 02660
9730
91030
4
026-03156
9728
91199
6
-
9731
91031
7.12 Cable length
The maximum cable length is dependent on the chosen communication format (RS232, RS422 or
RS485), the baud rate, and, to a lesser extent, on the cable type and the local electrical ‘noise’
level.
The above table shows the typical maximum lengths at the given baud rates, using the
recommended cable. If any problems of data corruption etc. are experienced, then a slower baud
rate should be used. Alternatively, a thicker or higher specification cable can be tried.
Communication format
Baud rate
Max. cable length
RS232 (PCIA Output)
9600
6.5 m (20 ft) or if using Cat 5 cable see
manufacturers recommendations
RS422
9600
1 km (3200 ft)
Analogue outputs - voltage
N/A
6.5 m (20 ft)