![Gibertini Crystal 100 Instruction Manual Download Page 47](http://html1.mh-extra.com/html/gibertini/crystal-100/crystal-100_instruction-manual_2221910047.webp)
Balanza Analitica CRYSTAL
ESPANOL
MANUAL DE INSTRUCCIONES - 60.50.630 9 revisión - Diciembre 2014
Pàgina 45 de 48
PRINTER protocol
Is activated by the “PRINT” pushbutton. The output string uses 10 characters for the measure
including sign and decimal point, the value is right justified and filled with spaces from the left. The
format of the output is:
D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B U U U CR LF
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
16
Do to D9
Measure
with sign & decimal point, right justified
B
Blank
Space between measure and unit
U
Unit
Measuring unit
CR
Carriage Return
ASCII code <CR>
LF
Line Feed
ASCII code <LF>
Note :
Data will be printed if and only if the wheight on pan is stable.
11.3. REMOTE CONTROL COMMANDS
The balance responds to the following commands:
Command
Function
B
Send data on display
C
Perform calibration
F
Stop continuous transmission of data on display
I
Start continuous transmission of data on display
T
Tare command
Notes:
Every remote control command must end with <CR>.
Line-feed characters are considered dont’care symbols