![Gibertini Crystal 100 Instruction Manual Download Page 46](http://html1.mh-extra.com/html/gibertini/crystal-100/crystal-100_instruction-manual_2221910046.webp)
Pàgina 44 de 48
MANUAL DE INSTRUCCIONES - 60.50.630 9 revisión - Diciembre 2014
CRYSTAL Protocol
The answer to any type of command is formatted with a measure field, a unit field and a status flag
field as follows:
D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B
U
U
U
B
F1 F2 CR LF
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19
D0 …D9
Measure
with sign & decimal point, right justified
B
Blank
Space between measure and unit
U
Unit
Measuring unit
F1, F2
Status
Flags indicating the status of the balance
CR
Carriage Return
ASCII code <CR>
LF
Line Feed
ASCII code <LF>
The status flags values of F1 are:
D
Transmitted data is valid
O
Balance in over-range
U
Balance in under-range
T
Tare in progress
C
Calibration in progress
Z
Zero acquire
I
Init system
The status flags values of F2 are:
S
Measure is stable
I
Measure is not stable
E
Measure is in error
A
Automatic calibration ackowleged
B
Calibration in progress
EURO protocol
The format of the euro protocol is:
D0 D1 D2 D3 D4 D5 D6 D7 D8 CR LF
1
2
3
4
5
6
7
8
9
10
11
D0 … D8
Measure
with sign & decimal point
CR
Carriage Return
ASCII code <CR>
LF
Line Feed
ASCII code <LF>