CRYSTAL Protocol 

 
The answer to any type of command is formatted with a measure field, a unit field and a status flag 
field as follows: 
 

D0  D1  D2  D3  D4  D5  D6  D7  D8  D9  B 

F1  F2  CR  LF 

10  11  12  13  14  15  16  17  18  19 

 

 
D0 …D9 

Measure 

with sign & decimal point, right justified 

Blank 

Space between measure and unit 

Unit 

Measuring unit 

F1, F2 

Status 

Flags indicating the status of the balance 

CR 

Carriage Return 

ASCII code <CR> 

LF 

Line Feed 

ASCII code <LF> 

 

The status flags values of F1 are: 

Transmitted data is valid 

Balance in over-range 

Balance in under-range 

Tare in progress 

Calibration in progress 

Zero acquire 

Init system 

 
The status flags values of F2 are: 

Measure is stable 

Measure is not stable 

Measure is in error 

Automatic calibration ackowleged 

Calibration in progress 

 
 

EURO protocol 

 
The format of the euro protocol is: 
 

D0  D1  D2  D3  D4  D5  D6  D7  D8  CR  LF 

10 

11 

 

D0 … D8 

Measure 

with sign & decimal point 

CR 

Carriage Return 

ASCII code <CR> 

LF 

Line Feed 

ASCII code <LF> 

 

Summary of Contents for Crystal 100

Page 1: ...N MANUAL MANUAL DE USUARIO ISTRUZIONI PER L USO INSTRUCTION MANUAL MANUAL DE USUARIO ISTRUZIONI PER L USO INSTRUCTION MANUAL Bilancia Analitica Serie Crystal MANUAL DE USUARIO ISTRUZIONI PER L USO IST...

Page 2: ...ting The manufacturer reserves the right to make changes to the product at any time without notice It is advisable to verify if there are any updates Tutti i diritti riservati E vietata qualsiasi form...

Page 3: ......

Page 4: ...IA CON FUNZIONE CONTA PEZZI 9 9 1 3 BILANCIA CON FUNZIONE DI CALCOLO PERCENTUALE 10 9 2 SET UP 11 9 2 1 SERIAL 12 9 2 1 1 BAUD 12 9 2 1 2 BYTE FOR 12 9 2 1 3 PROTOC 12 9 2 1 4 TRANS OPT 12 9 2 2 CALIB...

Page 5: ...STALLATA IN AMBIENTI ASCIUTTI PULITI E SENZA VIBRAZIONI NON DEVE ESSERE INSTALLATA IN ZONE A RISCHIO DI ESPLOSIONE NOTA In caso di impostazioni utente non valide la bilancia potrebbe non funzionare co...

Page 6: ...I SI 3 POSIZIONAMENTO Scegliere una posizione adatta considerando i seguenti criteri Lontano da vibrazioni e da correnti d aria Il basamento del supporto deve essere rigido La tensione della rete deve...

Page 7: ...pra indicata Cilindro imballo Livellare la bilancia ruotando i piedini regolabili sotto la parte anteriore del basamento FIG 3 Installare l anello di protezione NON presente nel modello 510g ed inseri...

Page 8: ...na 6 di 48 ISTRUZIONI PER L USO 60 50 630 Ed 9 Dicembre 2014 5 COMANDI DI UTILIZZO DELLA BILANCIA 1 2 4 5 7 6 3 FIG 3 Vista d assieme 1 Display 2 Tasto ON OFF 3 Tasto Tara 4 Tasto Mode 5 Tasto Stampa...

Page 9: ...secondo attivano la seconda solo i tasti MODE e ON OFF Lo specchietto seguente riassume le varie possibilit Funzionamento normale NOME TASTO ATTIVAZIONE DEL TASTO PER MENO DI UN SECONDO ATTIVAZIONE DE...

Page 10: ...tto men selezionato mentre il tasto TARA fa uscire dal men principale Quando si entra in un sotto men i tasti funzionano come sopra premendo ON OFF la selezione confermata premendo TARE la selezione r...

Page 11: ...rare istantaneamente lo stesso peso con due unit di misure differenti Tale funzione agisce su tutte le seconde unit di misura presenti in CRYSTAL Quando una seconda unit di misura selezionata essa sar...

Page 12: ...quisizione avranno come riferimento percentuale il campione precedentemente acquisito CRYSTAL pu misurare valori percentuali del 500 rispetto al campione purch non venga superata la portata massima de...

Page 13: ...48 9 2 SET UP 8 E 1 SERIAL BAUD PROTOC BYTE FOR 1200 2400 4800 9600 7 E 1 7 O 1 7 N 2 7 E 2 7 O 2 8 N 1 8 N 2 8 O 1 CRYSTAL PRINTER EURO TRANS OPT REQUIEST CONTINUE STABLE OPT STABLE OFF STABLE ON CA...

Page 14: ...tere dati ad una periferica che non in grado di inviare comandi alla bilancia la trasmisione continua si attiva all accensione in modo che ad ogni misura effettuata dalla bilancia segua la trasmission...

Page 15: ...ogni cambiamento con la funzione SET DEF Le scelte possibili sono 9 2 5 1 END i parametri non vengono memorizzati 9 2 5 2 STORE tutti i parametri modificati vengono memorizzati 9 3 CAL E Quando viene...

Page 16: ...minuti nessun comando seriale e stato inviato alla bilancia Negli ultimi 10 minuti non variato il valore letto sul display Se questa funzione termina correttamente il display mostrer in modo permanent...

Page 17: ...unque tipo di comando accettato formattata con un campo misura un campo unit e un campo di stato come segue D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B U U U B F1 F2 CR LF 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1...

Page 18: ...di uscita D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B U U U CR LF 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 D0 D9 Misura Con segno e punto decimale giustificata a destra B Blank Spazio tra misura ed unit U Unit Unit...

Page 19: ...UNTING FUNCTION 24 9 1 3 PERCENTAGE FUNCTION 25 9 2 SETUP 26 9 2 1 SERIAL 27 9 2 1 1 BAUD 27 9 2 1 2 BYTE FOR 27 9 2 1 3 PROTOC 27 9 2 1 4 TRANS OPT 27 9 2 2 CALIBR 27 9 2 2 1 CAL OFF 27 9 2 2 2 CAL O...

Page 20: ...ly pack matches the line voltage CRYSTAL MAY ONLY BE USED WITH THE POWER SUPPLY PACK SUPPLIED MUST ONLY BE OPERATED IN DRY CLEAN AND VIBRATION FREE ENVIRONMENT MUST NOT BE OPERATED IN AREAS WHERE THER...

Page 21: ...Internal Calibration NO YES NO YES YES YES 3 LOCATION Choose an appropriate location taking into account these criteria Vibrations and air drafts in the ambient must be avoided The supporting stand m...

Page 22: ...cket can be refastened to the balance by reolacing the knob locking bracket Level the balance using the spirit level on the rear and rotating the levelling feet under the front part of the basement Fi...

Page 23: ...MANUAL 60 50 630 9th issue December 2014 Page 21 of 45 5 OPERATING CONTROLS OF THE BALANCE 1 2 4 5 7 6 3 FIG 3 General view 1 Main Display 2 On Off Pushbutton 3 Tare Pushbutton 4 Mode Pushbutton 5 Pri...

Page 24: ...the keys are activated for less than 1 second the first function is activated when pressed for more the second function will be active The table below shows the operation capabilities Normal use KEY N...

Page 25: ...pear The ON OFF key activates the selected submenu while the TARE key exits the main menu When the submenu is entered the keys operate as above pressing ON OFF the selection is confirmed while pressin...

Page 26: ...e resolution of the balance total number of divisions is reduced so that the last digit always exhibits a 1 count increment When a second unit is selected the balance automatically stores the informat...

Page 27: ...play the the new reference value is acquired and the balance starts to display the weight as a percentage of the reference acquired Notes If the weight of the sample is too small acquisition is inhibi...

Page 28: ...2014 9 2 SETUP 8 E 1 SERIAL BAUD PROTOC BYTE FOR 1200 2400 4800 9600 7 E 1 7 O 1 7 N 2 7 E 2 7 O 2 8 N 1 8 N 2 8 O 1 CRYSTAL PRINTER EURO TRANS OPT REQUIEST CONTINUE STABLE OPT STABLE OFF STABLE ON CA...

Page 29: ...tocols refer to the section relative to the data transmission 9 2 1 4 TRANS OPT Selects the continous or on demand data transmission When the continous mode is selected the balance transmits a data at...

Page 30: ...e different parameters To mantain a custom configuration that will be active at power on the function SET DEF was implemented 9 2 5 1 STORE current settings are stored as user default 9 2 5 2 END user...

Page 31: ...m the last calibration the balance must be inactive for at least 10 minute before the calibration time no serial command was send to the balance The display shows CAL AUTO during the calibration proce...

Page 32: ...of command is formatted with a measure field a unit field and a status flag field as follows D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B U U U B F1 F2 CR LF 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 D0 D9...

Page 33: ...he format of the output is D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B U U U CR LF 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Do to D9 Measure with sign decimal point right justified B Blank Space between measure and...

Page 34: ...EZAS 39 9 1 3 FUNCI N PORCENTAJE 39 9 2 SETUP 40 9 2 1 SERIAL 41 9 2 1 1 VELOCIDAD 41 9 2 1 2 BYTE 41 9 2 1 3 PROTOC 41 9 2 1 4 TRANS OPT 41 9 2 2 CALIBRACION 41 9 2 2 1 CAL OFF 41 9 2 2 2 CAL ON 41 9...

Page 35: ...L ADAPTADOR SUMINISTRADO CON LA BALANZA LA BALANZA CRISTAL DEBE NICAMENTE UTILIZARSE EN UN AMBIENTE SECO LIMPIO Y LIBRE DE VIBRACIONES LA BALANZA NO DEBE UTILIZARSE EN REAS DONDE HAYA RIESGO DE EXPLOS...

Page 36: ...ANZA Elegir un lugar apropiado para instalar la balanza siguiendo los siguientes criterios Deben evitarse corrientes de aire y vibraciones El soporte de la balanza debe ser r gido La fuente de aliment...

Page 37: ...ue ser transportada Colocar el plato pieza de bloqueo Nivelar la balanza usando el nivel situado en la parte de atr s de la balanza y las patas regulables FIG 3 Encajar el embellecedor de aluminio No...

Page 38: ...CCIONES 60 50 630 9 revisi n Diciembre 2014 5 DESCRIPCION DE LOS CONTROLES DE LA BALANZA 1 2 4 5 7 6 3 FIG 3 1 Display 2 Pulsador ON OFF 3 Pulsador TARE 4 Pulsador MODE 5 Pulsador PRINT 6 Plato de pes...

Page 39: ...a tecla MODE es utilizada para controlar funciones adicionales de la balanza de este modo un mismo pulsador puede comprender varias funciones seg n como pulsemos una pulsaci n corta nos da lugar a una...

Page 40: ...ado en el submen operamos como sigue a continuaci n Pulsando la tecla confirmada selecci n es confirmada mientras que pulsando TARA la selecci n es dejada sin cambiar NOTA Para los modelos sin la cali...

Page 41: ...gramos pcs indistintamente para seleccionar el n mero de piezas nos situamos en la funci n pcs y hacemos una pulsaci n mantenida de Mode el display nos indicar 10 pcs dejamos de pulsar y pulsamos rep...

Page 42: ...P 8 E 1 SERIAL BAUD PROTOC BYTE FOR 1200 2400 4800 9600 7 E 1 7 O 1 7 N 2 7 E 2 7 O 2 8 N 1 8 N 2 8 O 1 CRYSTAL PRINTER EURO TRANS OPT REQUIEST CONTINUE STABLE OPT STABLE OFF STABLE ON CAL OFF CALIB S...

Page 43: ...Por una descripci n detallada de los diferentes protocolos acudir a la secci n de la transmisi n de los datos 9 2 1 4 TRANS OPT Seleccionar el modo de transmisi n de datos continua o requerida Cuando...

Page 44: ...configuraci n actuales son memorizada 9 2 5 2 END Los ajustes del usuarios son dejados sin cambiar por defecto 9 3 CALIBRACI N EXTERNA CAL E Una vez que la calibraci n externa se activa el display mue...

Page 45: ...brica Quitar de la alimentaci n principal Mantener pulsado la tecla ON OFF y encender la balanza Esperar hasta que el display nos muestre Cristal Soltar la tecla ON OFF El display nos mostrara SET FAC...

Page 46: ...dicating the status of the balance CR Carriage Return ASCII code CR LF Line Feed ASCII code LF The status flags values of F1 are D Transmitted data is valid O Balance in over range U Balance in under...

Page 47: ...14 15 16 Do to D9 Measure with sign decimal point right justified B Blank Space between measure and unit U Unit Measuring unit CR Carriage Return ASCII code CR LF Line Feed ASCII code LF Note Data wi...

Page 48: ...P gina 46 de 48 MANUAL DE INSTRUCCIONES 60 50 630 9 revisi n Diciembre 2014...

Reviews: