8 | Technical data
G 1400 series
16 / 20
B-H87.0.0X.DK2-2.1
G 1410
Measuring
range
Conductivity
0 .. 2000 S/cm
0.00 .. 20.00 mS/cm
0.0 .. 100.0 mS/cm
Specific re-
sistance
-
Salinity
0.0 .. 50.0 g/kg
TDS
0 .. 2000 mg/l
Temperature
-5.0 .. +105.0 °C (23.0 .. +221.0 °F)
– the conduc-
tivity measuring cells can be exposed temporarily
to temperatures of up to 100 °C and permanently
to temperatures of up to 80 °C.
Accuracy
Conductivity
± 0.5 % of measured value ± 0.5 % FS
Temperature
± 0.3 °C
Measuring cycle
approx. 10 measurements per second
Updating of the display approx. 2 times per second
Display
3-line segment LCD, additional symbols, illuminat-
ed (adjustable white, permanent illumination)
Additional functions
Min/Max/Hold
Adjustment
Offset and gradient correction - temperature,
Gradient correction - conductivity
Housing
Break-proof ABS housing
Protection rating IP65 / IP67
Dimensions
L*W*H [mm] and
weight
108 * 54 * 28 mm without measuring cell or kink
protection
180 g, incl. battery and measuring cell