8 | Technical data
G 1400 series
14 / 20
B-H87.0.0X.DK2-2.1
8
Technical data
_
_
G 1409
Measuring
range
Conductivity
0 .. 5000 µS/cm
EC
0 .. 5.000 EC (corresponds to mS/cm)
CF
0.00 .. 50.00 CF(corresponds to 10 x EC)
TDS
0 .. 2000 mg/l
Temperature
-5.0 .. +80.0 °C (23.0 .. +176.0 °F).
Accuracy
Conductivity
typ. ± 1 % of measured value ± 0.5 % FS
(0 .. 2000 µS/cm)
Temperature
± 0.3 °C
Measuring cycle
approx. 10 measurements per second
Updating of the display approx. 2 times per second
Display
3-line segment LCD, additional symbols, illuminat-
ed (adjustable white, permanent illumination)
Additional functions
Min/Max/Hold
Adjustment
Offset and gradient correction - temperature,
Gradient correction - conductivity
Housing
Break-proof ABS housing
Protection rating IP65 / IP67
Dimensions
L*W*H [mm] and
weight
108 * 54 * 28 mm without measuring cell or kink
protection
180 g, incl. battery and measuring cell
Operating conditions
-20 to 50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.)
Storage temperature
-20 to 70 °C