Temperature compensation
TCOR
OFF
Do not compensate conductivity measurement
NLF
Non-linear function for natural water in accordance
with EN 27888 (ISO 7888) Ground water, surface
water or drinking water
NACI
Compensation of weak NaCI solutions only in pure
and ultrapure water (G 1420 only)
L,N
Linear temperature compensation (G 1420 only)
_
Compensation coefficient
TL,N
(G 1420 only)
0
.
300
..
3
.
000
Temperature compensation coefficient in %/K
_
Reference temperature for temperature compensation
TREF
25
°
(
Reference temperature 25 °C or 77 °F
20
°
(
Reference temperature 20 °C or 68 °F
_
Shut-off time
POFF
OFF
No automatic shut-off
15
30
60
120
240
Automatic shut-off after a selected time in minutes,
during which no buttons have been pressed
_