E40.0.01.6C-08
Operating Manual EB-Logg 80CL
page 9 of 16
5.3 Operating possibilities at the device - Choose number of
EASYBus-Channels
Function
Procedure
Show and adjust
the measurement
channels, address-
able via EASYBus
Push button
and
at the same time.
The main display shows „S ch“. The secondary display shows
the number of channels, which are addressable via EASYBus.
Adjust via button
and
the desired number of chan-
nels (1..4).
Important: To change the number of channels, the logger-
display must show
‚Stop‘, so no data saved, and the logger
must not be connected to the EASYBus.
Press Button
again to confirm your adjustments and to
leave the menu..
Note:
If no key is pressed for more than 2 min, the menu is
quit without saving.
This setting allows the logger, to be a substitute for outdated loggers in existing EASY-
Bus installations.
The table below shows
the channels, depending on the selection of ‚S ch‘, which are
addressable via EASYBus.
S ch
Addressable measuring channels
Substitute for
4
Humidity
Temperature
Air pressure
calculated variable (according selection)
3
Humidity
Temperature
Air pressure
2
Humidity
Temperature
… 24RFT
… 24RFT-E
1
Humidity
… 40RF
Independent from the number of selected measuring channels, the EB-Logg 80CL
measures all 4 channels and displays them. But the logger only records the active
channels for EASYBus. Therefore only for these channels, data can be read out.
SET
1
3
2
3
SET
1