10. Frequently asked questions [FAQ]
Problem:
The logger cannot be stopped via keys.
Possible reason:
“Logger stop via buttons disabled” has been activated at the configu-
ration of the logger.
Remedy:
Deactivate this function by means of GSOFT
40K (at tab „settings“).
Problem:
The logger cannot be stopped both via keys and GSOFT 40K.
Possible reason:
“Do not stop active logger” has been activated at the configuration of
the logger.
Remedy:
Deactivate this function by means of EASYBus-configurator.
Problem:
The alarm settings or the sea-level correction cannot be
changed.
Possible reason:
The logger has still saved data.
Remedy:
Stop logger as recommended by the software (delete data).
Problem:
The desired calculated variable cannot be changed.
Possible reason:
The logger has still saved data.
Remedy:
Stop logger as recommended by the software (delete data).
Problem:
A error message is displayed.
Error message
Possible reason
procedure / remedy
measuring range
exceeded
Measuring range of the chan-
nel has been exceeded.
Operate logger within its specifica-
tions.
humidity: sensor bedewed
Twist filter cap carefully off and let
sensor dry.
Sensor defective
Send the logger to the manufactur-
er for repair.
measuring range
under-run
Measuring range of this chan-
nel is under-run.
Operate logger within its specifica-
tions.
Sensor defective
Send the logger to the manufactur-
er for repair.
system error
Recording error
Reset the error via GSOFT 40K.
If the error message is displayed
furthermore send the logger to the
manufacturer for repair.
System error
Send the logger to the manufactur-
er for repair.
sensor error
Humidity / temperature sensor
has moved out of its socket
Check the position of humidity /
temperature sensor.
Sensor defective
Send the logger to the manufactur-
er for repair.
calculation error
Error message at a variable
needed for the calculation
(humidity / temperature / pres-
sure)
Remedy the reason for error at this
channel.