E40.0.01.6C-08 

Operating Manual EB-Logg 80CL 

page 10 of 16 

5.4  Operating possibilities via software GSOFT 40K 

You can easily operate the logger via the software GSOFT 40K (> V7.14). This software 
provides a variety of functions to operate the logger. For example: 

•  start and stop the logger 
•  read out the logger data; illustrate them graphically and save them 
•  adjust the alarm boundaries and alarm delay 

*

1

 

•  activate the sea-level correction and enter the actual sea level 

*

1

 

Description: 

The measured barometric air pressure can be converted to the air 
pressure at sea level if the actual height above sea level is entered.    

•  activate the function: “logger stop via buttons disabled“ 

*

1

 Please note:  

The alarm settings and the sea-level correction can be only 
changed if the logger is stopped (= logger without logger data). 

5.5  Operating possibilities via software EASYBus-Configurator 

The software EASYBus-Configurator (> V2.0) is needed to change the configuration 
settings.  
With that software the following adjustment are additionally possible.  

•  Selection of the unit of the calculated variables

*

2

  

•  Offset and slope adjustment for each measured variable 

*

2

 

•  Activation of the function: “do not stop active logger” 

*

2

 Please note:  

This can be only changed if the logger is stopped (=logger without 
logger data) 

6. Battery life time and recording period 

 

Measuring cycle:

 

4 sec 

3 min 

15 min 

5 h 

 

Recording period:

 

11.5 days 

521 days 

7.1 years 

142 years 

 

Battery life time:

 

-- 

-- 

ca. 5 years 

-- 

 
 

 

Please note:

 

Short measuring cycles seriously reduce the battery life time!  
 
In this case you should consider not unplugging the 

EASYBus

-

interface.   
Then the logger is supplied via the interface. 
The internal battery is spared. 

 

 

 

 

Summary of Contents for GREISINGER EB-Logg 80CL

Page 1: ...ik GmbH Hans Sachs Str 26 93128 Regenstauf GERMANY Tel 49 9402 9383 0 info greisinger de www greisinger de from from Version V2 7 Operating Manual EB Logg 80CL EB Logg 80CL E Data logger for humidity...

Page 2: ...Operating possibilities via software EASYBus Configurator 10 6 Battery life time and recording period 10 7 Operating displays 11 7 1 Measured value display 11 7 2 Logger status messages 11 7 3 System...

Page 3: ...ge This device must not be used at a patient for diagnostic or other medical purpose The liability and warranty of the manufacturer for damages and consequential damages are voided with misuse disrega...

Page 4: ...only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has to...

Page 5: ...tings 3 2 Notes on the logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a communication with t...

Page 6: ...logger is recording BAT signals weak battery see chapter 7 AL signals an alarm state at minimum one channel SL signals an active sea level correction for the air pressure 4 2 Function of the buttons b...

Page 7: ...the display el ement MAX Note You can immediately switch to the next measuring channel pushing the button again Delete max value Push button for about 5 seconds CLr is displayed and the max values of...

Page 8: ...tart recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with the arrow keys button or and verify with Note If no key is p...

Page 9: ...rm your adjustments and to leave the menu Note If no key is pressed for more than 2 min the menu is quit without saving This setting allows the logger to be a substitute for outdated loggers in existi...

Page 10: ...an be only changed if the logger is stopped logger without logger data 5 5 Operating possibilities via software EASYBus Configurator The software EASYBus Configurator V2 0 is needed to change the conf...

Page 11: ...row AL signals an alarm state at minimum one channel The arrow SL signals an active sea level correction 7 2 Logger status messages At stand alone operation no EASYBus connected logger status messages...

Page 12: ...repair 7 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured value is below the selected min alarm lim...

Page 13: ...RH Air pressure 1 0 mbar typ 2 5 mbar max Display two 4 digit LCD displays about 7 and 5 8 mm high Interface EASYBus Busload equals to 2 EASYBus elements Battery 1x fixed lithium cell Li SOCl2 3 6V 1...

Page 14: ...lowing European di rectives 2011 65 EU RoHS 2014 30 EU EMV directives Applied harmonized standards EN 50581 2012 EN 61326 1 2013 emissions level class B emi immunity according to table 1 additional er...

Page 15: ...delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceeded Measuring range of the chan nel has been exceeded Operate logger within its sp...

Page 16: ...es a General connection diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connec...

Reviews: