HD51.3D…
- 25 -
V2.0
Units of measurement:
Command
Reply
Description
CGUVn
&|
Sets measuring unit of wind speed:
m/s if n=1
cm/s if n=2
km/h if n=3
knot if n=4
mph if n=5
Default
: m/s (n=1)
RGUV
n|
Reads the wind speed measuring unit set
in the instrument
CGUTn
&|
Sets temperature measuring unit:
°C if n=1
°F if n=2
Default
: °C (n=1)
RGUT
n|
Reads the temperature measuring unit set
in the instrument
CGUPn
&|
Sets pressure measuring unit:
mbar if n=1
[Note:1
mbar=1
hPa]
mmHg if n=2
inchHg if n=3
mmH
2
O if n=4
inchH
2
O if n=5
atm if n=6
Default
: mbar (n=1)
RGUP
n|
Reads the pressure measuring unit set in
the instrument
General parameters:
Command
Reply
Description
CGHn
&|
Enables/disables heating:
Disables if n=0
Enables if n=1
Default
: Enabled (n=1)
RGH
n|
Reads heating enabling state set in the in-
strument
CWCnnnn
&|
Sets wind speed threshold to nnnn value (in
hundredths of m/s)
Value should range within 0 and 100 hun-
dredths of m/s (= 0…1 m/s)
Default
: 20 (= 0.2 m/s)
(see
Note 3
)
RWC
& nnnn|
Reads the wind speed threshold value set in
the instrument (in hundredths of m/s)
CWaLnnn
&|
Sets time interval for the calculation of aver-
age speed and average direction to nnn value
Value should range within 1 and 600 s
Default
: 1 s