![GHM DeltaOHM HD51.3D Series Operating Manual Download Page 22](http://html1.mh-extra.com/html/ghm/deltaohm-hd51-3d-series/deltaohm-hd51-3d-series_operating-manual_2220418022.webp)
HD51.3D…
- 22 -
V2.0
1
1
2
2
4
4
3
3
N
x
ws
y
ws
x
ws
y
ws
Housing in technopolymer
Housing in aluminium alloy
Digit 0 indicates that no errors are present in the ultrasonic transducers.
The second digit of the error code indicates the type of error:
0
= no errors;
5
= transducer broken, electric interruption, path obstruction;
Other
= codes re-
served to technical service.
2.
Second number
= heating status:
0
= off,
1
= housing heating is on,
2
= housing heat-
ing and transducers heating are both on.
3.
Third number
= number of invalid measurements.
Example
: if
25 0 2
appears in correspondence to the error condition in the data string sent by
the instrument, it means that an error occurred in the transducer number 2, that the heating is
turned off and that two measurements have been rejected due to the error occurrence.