Note 1
: after sending the CUMn command, the instrument remains in configuration mode.
Power cycle the instrument to activate the set operating mode.
Parameters for RS232 and RS485 ASCII proprietary modes:
Command
Reply
Description
CU1Ac
&|
Sets the address for RS485 ASCII
proprietary mode to c value
The address is an alphanumeric character
ranging within 0…9, a…z, A…Z
Default
: 0
RU1A
& c|
Reads the address for RS485 ASCII pro-
prietary mode set in the instrument
CU1Bn
&|
Sets Baud Rate for RS485 ASCII
proprietary mode to:
9600 if n=3
19200 if n=4
38400 if n=5
57600 if n=6
115200 if n=7
Default
: 115200 (n=7)
RU1B
& n|
Reads Baud Rate setting for RS485 ASCII
proprietary mode
CU2Bn
&|
Sets Baud Rate for RS232 ASCII
proprietary mode to:
9600 if n=3
19200 if n=4
38400 if n=5
57600 if n=6
115200 if n=7
Default
: 115200 (n=7)
RU2B
& n|
Reads Baud Rate setting for RS232 ASCII
proprietary mode
CU2In
&|
Sets interface for ASCII mode to:
RS232 if n=0
RS485 if n=1
RS422 if n=2
Default
: RS485 (n=1)
RU2I
& n|
Reads interface setting for ASCII mode
CU2Mn
&|
Sets parity and stop bits for ASCII mode
to:
8N1 if n=0 [No parity, 1 stop bit]
8N2 if n=1 [No parity, 2 stop bits]
8E1 if n=2 [Even parity, 1 stop bit]
8E2 if n=3 [Even parity, 2 stop bits]
8O1 if n=4 [Odd parity, 1 stop bit]
8O2 if n=5 [Odd parity, 2 stop bits]
The number of data bits is fixed to 8
Default
: 8N2 (n=1)