![GHM DeltaOHM HD37AB1347 Operating Manual Download Page 46](http://html.mh-extra.com/html/ghm/deltaohm-hd37ab1347/deltaohm-hd37ab1347_operating-manual_2220417046.webp)
HD37AB1347
- 46 -
V3.0
L
OGGING
F
UNCTION
The
Logging
function allows recording of the measurements registered by the probe
connected to the inputs. The time interval between two consecutive measurements
can be set from 15 seconds to 1 hour. The logging starts by pressing the
MEM
key
and ends by pressing the same key again: the data memorized in this way form a
continuous block of data.
See the description of the menu items on chapter
“
MAIN MENU
”.
If the automatic turning off option between two recordings (see par.
AutoPowerOff–
AutoPowerOff Mode
)
is enabled, upon pressing the
MEM
key the instrument logs the
first data and turns off. 1 minute before the next logging instant, it turns on again to
capture the new sample, and then turns off.
The data stored in the memory can be transferred to a PC through the DeltaLog10
software. During data transfer the display shows the message DUMP; to stop the data
transfer press ESC on the instrument or on the PC.
C
LEARING THE MEMORY
To clear the memory, use the Erase Log function. The instrument starts clearing the
internal memory; at the end of the operation, it goes back to normal display.
NOTES
:
•
Data transfer does not cause the memory to be erased: The operation can be
repeated as many times as required.
•
The stored data remain in the memory independently of the battery charge
conditions.
•
In order to print the data to a parallel interface printer, you must use a parallel-
serial adaptor (not supplied).
•
The direct connection between instrument and printer via a USB
connector does not work.
•
Some keys are disabled during
logging
. The following keys are enabled:
MEM
,
MENU
,
ENTER
and
ESC
.
•
Pressing the
MEM
and
MENU
keys has no effect on the logged data if these keys
are pressed
after
starting the recording, otherwise the following is valid.
P
RINT
F
UNCTION
Press
ENTER
to send the measured data directly to the RS232 or USB port, in real
time. The printed data units of measurements are the same as those used on the
display. The function is started by pressing
ENTER
.
The time interval between two consecutive prints can be set from 15 second to 1 hour
(please see the
Print interval
menu
item). If the print interval is equal to 0, by
pressing
ENTER
a single data is sent to the connected device. If the print interval is
higher than 0, the data transfer continues until the operator stops it by pressing
ENTER
again.
NOTE: When setting the baud-rate, check the printer speed.