![GHM DeltaOHM HD37AB1347 Operating Manual Download Page 45](http://html.mh-extra.com/html/ghm/deltaohm-hd37ab1347/deltaohm-hd37ab1347_operating-manual_2220417045.webp)
HD37AB1347
- 45 -
V3.0
Command
Response
Description
P0
&
Ping (locks the instrument keyboard for 70 seconds)
P1
&
Unlocks the instrument keyboard
S0
G0
Model HD21ABE17
Instrument model
G1
M=Indoor Air Quality
Model description
G2
SN=12345678
Instrument serial number
G3
Firm.Ver.=01.00
Firmware version
G4
Firm.Date=2010/02/10
Firmware date
G5
cal 2010/02/10 10:30:00
Calibration date and time
C1
RH-T probe type, serial number, calibration date
C2
CO-CO
2
probe type, serial number, calibration date
GC
Print instrument’s heading
GB
ID=0000000000000000
User code (set with T2xxxxxxxxxxxxxxxx)
HA
Print the current data measurement
LR
Print the instrument memory map
KInn
Print the information of Loggging nn
KRaaaa
Print the recorded data in page aaaa
KE
&
Stop the data download
LE
&
Erase stored data
K1
&
Immediate printing of data
K0
&
Stop printing data
K4
&
Start logging data
K5
&
Stop logging data
KP
&
Auto-power-off function = ENABLE
KQ
&
Auto-power-off function = DISABLE
WC0
&
Setting SELF off
WC1
&
Setting SELF on
RA
Sample print = 0sec
Reading of PRINT interval set and label of the
measurements
RL
Sample log = 30sec
Reading of LOG interval set
WA#
&
Setting PRINT interval.
# is a hexadecimal number 0…D that represents the
position of the interval in the list 0, 1, 5, 10, …, 3600
seconds.
WL#
&
Setting LOG interval.
# is a hexadecimal number 1…D that represents the
position of the interval in the list 15, …, 3600 seconds.
S
TORING AND
T
RANSFERING
D
ATA TO A
PC
The
HD37AB1347
instrument can be connected to a PC via an RS232C serial port or
USB port, and exchange data and information through the DeltaLog10 software
running in a Windows operating environment. It is possible to print the measured
values on the HD40.1 printer (
ENTER
key), or store them in the internal memory
using the
Logging
function (MEM key). If necessary, the data stored in the memory
can be transferred later to a PC.