background image

9

8850-2 Conductivity/Resistivity Transmitter

Parts Per Million (PPM) Factor

This feature is only applicable when PPM display units are selected. The programmable PPM Factor is adjustable from 0.01 to 3.00 
(factory default = 2.00).

Determine the best PPM Factor for a process solution by calculating the solution's conductivity (

μ

S) and the percent of total dissolved 

solids (PPM).

Temperature Coef

fi

 cient

Conductivity measurement is highly dependent on temperature. Temperature dependence is expressed as the relative change per °C, 
commonly known as percent/°C change from 25 °C, or slope of the solution.

Slopes can vary signi

fi

 cantly depending on process solution type. The factory default temperature compensation factor is 2.00%/°C. 

Process solutions may require adjustment for maximum accuracy. To determine the optimum temperature compensation factor for a 
process:

 1.

  Disable the 8850 temperature compensation % factor by entering 0.00.

 2. 

  Heat the sample solution close to the maximum process temperature. Place sensor in the sample solution allowing several 

minutes for stabilization. Access the VIEW menu and record the displayed temperature and conductivity values in the spaces 
provided:

 

Displayed temperature: 

T1 = _______ °C

 

Displayed conductivity: 

C1 = _______ 

μ

S

 3.

  Cool the sample solution close to the minimum process temperature. Place electrode in the sample solution allowing several 

minutes for stabilization. Record the displayed temperature and conductivity values.

Displayed temperature: 

T2 = _______ °C

Displayed conductivity: 

C2 = _______ 

μ

S

(A 10% change in conductivity between steps 2 and 3 is recommended.)

 4.

  Substitute the recorded readings (steps 2 and 3) into the following formula:

TC Slope =

100 × (C1 - C2)

(C2 × (T1 - 25)) - (C1 × (T2 - 25))

Example: A sample solution has a conductivity of 205 

μ

S @ 48 °C. After cooling the solution, the conductivity was measured at 

150 

μ

S @ 23 °C (C1 = 205, T1 = 48, C2 = 150, T2 = 23).

The TC is calculated as follows:

 

5.

  Enter 1.42 into the Temperature Comp % function in the OPTIONS menu.

TC Slope =

100 × (205 - 150)

=

5500

= 1.42% / °C

(150 × (48 - 25)) - (205 × (23 - 25))

3860

Do not use this procedure if the process solution is from 0.055 

μ

S to 0.1 

μ

S (18 M

Ω

 to 10 M

Ω

). An internal pure water 

curve is used for these ranges. The factory default setting of 2.00% per °C should be used.

PPM Factor =

Solution conductivity (

μ

S/cm)

Total dissolved solids (PPM)

TDS (PPM) =

Solution conductivity (

μ

S/cm)

PPM Factor

PPM Factor =

400 

μ

S/cm

= 2.00

200 PPM

Example: Solution conductivity = 400 

μ

S/cm, TDS = 200 PPM (mg/L)

Summary of Contents for Signet ProcessPro 8850-2

Page 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Page 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Page 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Page 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Page 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Page 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Page 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Page 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Page 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Page 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...

Page 11: ...11 8850 2 Conductivity Resistivity Transmitter Notes...

Page 12: ...085 CR Sanitary 0 01 SS 1 to 1 1 2 in 3 2819 S1C 159 000 087 CR Sanitary 0 01 SS 1 to 1 1 2 in Certified 3 2819 S2 159 000 086 CR Sanitary 0 01 SS 2 in 3 2819 S2C 159 000 088 CR Sanitary 0 01 SS 2 in...

Reviews: