8850-2 Conductivity/Resistivity Transmitter

Options Menu

Display

(Factory settings shown)

Description

Contrast:

3 >

Adjust the LCD contrast for best viewing. A setting of 1 is lowest contrast, 5 is highest.  Select 
lower contrast if the display is in warmer ambient surroundings.

Cond Decimal:

****.* >

Set the decimal to the best resolution for your application. The display will automatically scale 
up to this restriction. Select *****., ****.*, ***.** **.*** or *.****.

Averaging

Off >

OFF provides the quickest output response to changes in temperature. Longer averaging 
period produces more stable display and output response.

Temp Display

°C >

Select temperature units: ºC or ºF.

Temperature Comp %:

2.00 >

Set the percent change in Conductivity caused by a 1 ºC change in temperature. May be from 
0.00 to 10.00.

Loop Adjust

4.00 mA 

>

Adjust the minimum and maximum current output. Use this setting to match the system output 
to any external device. The display value represents the precise current output.
Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 mA

Loop Adjust

20.00 mA 

>

Test Loop:

 

>

Press 

 or 

 keys to manually order any output current value from 3.6 mA to 21.00 mA to test 

current loop output.

Test Relay 1:

 

>

Press 

 or 

 keys to manually toggle the state of Relay 1.

Test Relay 2:

 

>

Press 

 or 

 keys to manually toggle the state of Relay 2.

Summary of Contents for Signet ProcessPro 8850-2

Page 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Page 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Page 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Page 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Page 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Page 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Page 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Page 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Page 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Page 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...

Page 11: ...11 8850 2 Conductivity Resistivity Transmitter Notes...

Page 12: ...085 CR Sanitary 0 01 SS 1 to 1 1 2 in 3 2819 S1C 159 000 087 CR Sanitary 0 01 SS 1 to 1 1 2 in Certified 3 2819 S2 159 000 086 CR Sanitary 0 01 SS 2 in 3 2819 S2C 159 000 088 CR Sanitary 0 01 SS 2 in...

Reviews: