background image

8850-2 Conductivity/Resistivity Transmitter 

Calibrate Menu

Display

(Factory settings shown)

Description

Cell Constant

Standard >

Select CUSTOM only if you are connecting a certi

fi

 ed conductivity sensor. Select STANDARD 

for all other systems.

Cell: Standard

1 >

For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.

Cell: Custom

1.0000 >

For CUSTOM sensors: Enter the precise cell constant from the certi

fi

 cate provided with your 

sensor, or from the information label on the sensor.

Cond Units

μ

S >

Select from these options: 

μ

S, mS, k

Ω

, M

Ω

 or PPM.

PPM Factor

2.00 >

If the Units selection is PPM, set the ratio of Total Dissolved Solids to 

μ

S. The maximum 

allowable is 3 

μ

S = 1 PPM. The TDS factor can only be set in PPM. See page 9.

Set:

Temperature >

Provides a maximum 25 ºC offset to match temperature measurement to external reference.

Set:

Conductivity >

For use with STANDARD sensors: Perform this single-point wet calibration for most accurate 
results. Not required for CUSTOM sensors except for periodic system accuracy con

fi

 rmation. 

Enter all zeroes to restore factory calibration to TEMP and COND settings.

Loop Source:

Cond >

Select the input (Conductivity or Temperature) source to be associated with the current
loop output.

Loop Range: 

μ

S

0.0000  ›  100.00  >

Select the minimum and maximum values for the 4 to 20 mA current loop output. Be sure to 
modify this setting if you change the Cond. Units.

Relay1 Source:

Cond >

Select Conductivity or Temperature as the source for relay 1.

Relay1 Mode:

Low >

Select the mode of operation for relay 1: High, Low or Proportional Pulse.
The relay may be disabled (Off) if not in use.

Relay1 Setpnt:

10.0000 

μ

S >

In Low or High Mode, the relay will be energized when the Conductivity reaches this value. Be 

sure to modify this setting if you change the Cond. Units.

Relay1 Hys:

0.5000 

μ

S >

In Low or High mode, the relay will be de-energized at Setpoint ± Hysteresis, depending on 
High or Low Setpoint selection.

Relay1 Range: 

μ

S

10.0000  ›  40.0000  >

If Relay 1 is in PULSE mode, set the start and end point of the range and also set the 
maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.) Be sure to 
modify this setting if you change the Cond. Units.

Relay1 PlsRate:

120 Pulses/min 

>

The combined Relay1 Range and Relay1 Pulse Rate settings shown here indicate:
"Start pulsing when the Conductivity value is 10 

μ

S and increase the pulse rate up to the 

maximum of 120 pulses per minute when the Conductivity value reaches 40 

μ

S." The 8850 

will accept any value from 0 to 400.

Last CAL:

02-10-09 >

Use this “note pad” to record important dates, such as annual recerti

fi

 cation or scheduled 

maintenance.

Settings repeat for Relay 2.

Summary of Contents for Signet ProcessPro 8850-2

Page 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Page 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Page 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Page 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Page 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Page 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Page 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Page 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Page 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Page 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...

Page 11: ...11 8850 2 Conductivity Resistivity Transmitter Notes...

Page 12: ...085 CR Sanitary 0 01 SS 1 to 1 1 2 in 3 2819 S1C 159 000 087 CR Sanitary 0 01 SS 1 to 1 1 2 in Certified 3 2819 S2 159 000 086 CR Sanitary 0 01 SS 2 in 3 2819 S2C 159 000 088 CR Sanitary 0 01 SS 2 in...

Reviews: