9
Signet 2580 FlowtraMag Meter
Digital (S
3
L) Output (Compatible with 8900 Multi-Parameter Controller, 9900 and 9950 Transmitter)
•
To select S
3
L, use Bluetooth
®
App.
•
24 VDC power at a minimum of 1 amp is always be connected to the 2580.
•
The 8900 will display 0 (Zero) fl ow rate during periods of reverse fl ow
The 9900 and 9950 will display negative numbers to indicate reverse fl ow
•
The maximum cable length from the 2580 to the 8900 or 9900 depends on the 8900 or 9900 confi guration.
Refer to the 8900, 9900 or 9950 manual for complete information.
2580 Wiring to Signet 9900
2580 Wiring to Signet 9950
Refer to the 0486 Profi bus Concentrator manual for
Frequency wiring and programing instructions.
Wiring with Digital (S
3
L) Output
PWR–
9900
Power
Input
LOOP–
LOOP+
PWR+
9900 S
3
L
Terminal
DA
TA
GND
V+
SHLD
9900
Transmitter
S3L/Freq
24 VDC @ 1A min.
Power Supply
–
+
2580
Ground (White)
Data (Red)
Output cable
Black
Brown
Shield
Blue not used
Not used
2580
Power cable
Black
Red
Shield (EMC enhancement)
9950
Digital (S
3
L)
+V
FREQ
DATA
GND
+V
FREQ
DATA
GND
Loop1+
Loop1–
Loop2+
Loop2–
PWR+
PWR–
CH2
CH1
DC POWER
+V
FREQ
DATA
GND
X
No connection
X
No connection
2580
Output cable
Black
Brown
Shield
Blue not used
Not used
Ground (White)
Data (Red)
+V
FREQ
DATA
GND
9950
Frequency
+V
FREQ
DATA
GND
+V
FREQ
DATA
GND
Loop1+
Loop1–
Loop2+
Loop2–
PWR+
PWR–
CH2
CH1
DC POWER
X
No connection
X
No connection
2580
Output cable
Black
Brown
Shield
Blue not used
Not used
Ground (White)
Frequency (Red)
GF Confi g Tool App
set to S
3
L
GF Confi g Tool App set to S
3
L