3.2 Calibration Menu, Frequency Inputs
Legend
= Press Keypad
= Intelek-Pro
Display
= Decision
Block
= Calibration
procedure
1 8 8 8 8
X X X X X X X X
Table 1
Select upper display time-
base:
1= flow displayed in seconds
2= flow displayed in minutes
3= flow displayed in hours
4= flow displayed in days
Note: Accumulator not
affected by timebase.
"1" is displayed only when
previously selected.
"1" flashes when deselected.
K-factor Setting
MOD
Security
Code?
Yes
Unlock security
code using:
ENTER
Modify?
No
Yes
Select
next
item?
Yes
FLOW
Press to
set K-factor from
.0001 to 19999.
1 9 9 9 9
K - F A C T O R
MOD
Normal operation
ENTER
Recall original value
MOD
No
CAL
No
CAL
Decimal set
in next step
ENTER
Modify?
No
Yes
Select
next
item?
Yes
CAL
FLOW
Press to
select decimal
location, 0-4 plcs.
1 8 8 8 8
K - f D P = 0
MOD
Normal operation
Recall original value
MOD
No
ENTER
Modify?
No
Yes
Select
next
item?
Yes
CAL
FLOW
Press to select
upper display
timebase (1-4)
2
T I M E B A S E
MOD
Normal operation
Recall original value
MOD
No
Go to
A
See Table 1
3.3 Calibration Menu, Analog Inputs
4 to 20 mA option illustrated
Note:
The unit's software
recognizes which input card/
configuration is inserted,
therefore displaying the
corresponding calibration
value.
MOD
Security
Code?
Yes
Unlock security
code using:
ENTER
No
CAL
ENTER
Modify?
No
Yes
Select
next
item?
Yes
CAL
FLOW
Press to select
operation range,
(4-20 or 0-20 mA)
MOD
Normal operation
4
m A = 0 o r 4
Recall original value
MOD
No
12