Repair and adjustment
22
Service instructions
5015108-00
Edition 0403
Differential pressure gauge for dryer
To calibrate the differential pressure gauge for the dryer, proceed as follows.
1. Connect a PC with the CS 1000 program installed to the disinfector.
2. Close the machine doors.
3. Connect an external differential pressure gauge in parallel with the existing gauge on the
machine. Zero the external differential pressure gauge.
4. Start CS 1000 and choose Settings/Calibration/Analog Input Automatic...
5. Click Dryer pressure (A) and then Calibrate (B).
A
B
V1573
V1563
Summary of Contents for 8666
Page 1: ...SERVICE INSTRUCTIONS Washer disinfector Getinge 8666 8668 Mfg no SEV0320078 5015108 00 ...
Page 2: ......
Page 4: ......
Page 6: ...Safetyregulations 2 Serviceinstructions 5015108 00 Edition 0403 ...
Page 8: ......
Page 10: ...Technicaldata 2 Serviceinstructions 5015108 00 Edition 0403 ...
Page 12: ...Technicaldata 4 Serviceinstructions 5015108 00 Edition 0403 Alternativeconnectionarrangements ...
Page 14: ......
Page 24: ......
Page 26: ......
Page 33: ......
Page 37: ......
Page 39: ......
Page 47: ......
Page 49: ...Repairandadjustment 2 Serviceinstructions 5015108 00 Edition 0403 ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......