Repair and adjustment
15
Service instructions
5015108-00
Edition 0403
15. Set 20 mA on the conductivity meter with
Y
and
Z
.
16. Enter 200 (A) in CS 1000 and click Set High.
17. Check that the value (B) rises to 200.
18. Log off CS 1000.
19. Choose Output with
Y
and
Z
.
20. Press
[
. Sel. Type appears.
21. Choose LIN with
Y
and
Z
.
22. Press
[
.
23. Press
Y
and
Z
together several times until you are logged out of the conductivity meter.
A
B
V1566
Summary of Contents for 8666
Page 1: ...SERVICE INSTRUCTIONS Washer disinfector Getinge 8666 8668 Mfg no SEV0320078 5015108 00 ...
Page 2: ......
Page 4: ......
Page 6: ...Safetyregulations 2 Serviceinstructions 5015108 00 Edition 0403 ...
Page 8: ......
Page 10: ...Technicaldata 2 Serviceinstructions 5015108 00 Edition 0403 ...
Page 12: ...Technicaldata 4 Serviceinstructions 5015108 00 Edition 0403 Alternativeconnectionarrangements ...
Page 14: ......
Page 24: ......
Page 26: ......
Page 33: ......
Page 37: ......
Page 39: ......
Page 47: ......
Page 49: ...Repairandadjustment 2 Serviceinstructions 5015108 00 Edition 0403 ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......