Alarm Code Alarm Cause Solution
Sterilization failed! (1)adjust the tightness of door seal may behave leakage.(2)Replace the door seal.
E 04
E 05
Pressure can not
be exhausted!
(1)check the opened-solenoid valve(Ev1) and drain-solenoid valve(Ev5),take out and clean
the s
opened
pool.(2)check three-way piece to used-water tank,open and clean it.(3)check the
-Solenoid valve(Ev1) and drain-solenoid valve(Ev5) power is DC 24V.
E 06
Door is opened in
the working!
(1)check the door hook and
door signal switch is good? To maintain contact
contact point
between them.(2)open the door,check leading point of electronic lock is retracted?
Replace the electronic lock.
E 07
Operation over
time!
(1)adjust the tightness of door seal may behave leakage.(2)Replace the door seal.
(3)check the opened-solenoid valve(Ev1) and drain-solenoid valve(Ev5)and vacuum-solenoid
valve(Ev4),take out
and clean spool.(Judgment of leakage Method: mist of water
Drop on silicone tube and hot).
(4)check the heating bar,Judgment Methods: the
Ω ≈
measurement resultof resistance(
)
68
~73
(750W);50~55
(1000W).if measurement result of resistance(
)
,it is broken!
Ω
Ω
Ω ≈
∞
Ω
(5)
measurement result of resistance(
)
30
check the heating coil,Judgment Method:
Ω ≈
~37
(1500W).if measurement result of resistance(
)
,it is broken!and check
Ω
Ω ≈
∞
Ω
electronic emperature Relay of heating coil,
t
poor contact point between the connectors
and jacks.
(6)check the steam generator is blocked?
Judgment Methods: the silicone between
the water pump and water pump
-solenoid valve is be bloated.
(7)check water pump is be broken?
Judgment Methods: Very loud sound and water in silicone
have move forward?
E 09 Over temperature!
Measure:temperature sensor resistance value(
)
1000+(Ambient temperature
*3.8)
Ω ≈
℃
(1)if measurement result is OK,check poor contact point between the sensor and the
control board.(2)if measurement result is No,replacing the temperature sensor.
E 08 Over pressure!
check the opened-solenoid valve(Ev1) is be blocked? open the opened-solenoid valve
(Ev1),take out and clean s
pool.
Page 56 of 58
Summary of Contents for JN-23
Page 1: ...EURO CERT 1128 0197 English JN 23 MODEL SteamSterilizer InstructionManual CLASS B and S ...
Page 4: ...Page 03 of 58 ...
Page 5: ...Page 04 of 58 ...
Page 7: ...Page 06 of 58 ...
Page 8: ...Page 07 of 58 ...
Page 10: ...Page 09 of 58 JN 23 overall space required ...
Page 12: ...Page 11 of 58 ...
Page 13: ...Page 12 of 58 ...
Page 14: ...Page 13 of 58 ...
Page 40: ...Page 39 of 58 ...
Page 41: ...Fig 6 2 1 Fig 6 2 2 Fig 6 2 3 Page 40 of 58 ...
Page 42: ...Fig 6 2 4 Fig 6 2 5 Fig 6 2 6 Chapter 4 3 2 Chapter 4 3 2 Page 41 of 58 ...
Page 43: ...6 2 1 Periodic Maintenance Page 42 of 58 ...
Page 44: ...6 3 Extraordinary Maintenance Page 43 of 58 ...
Page 51: ...Page 50 of 58 ...
Page 52: ...134 04min 15min 3 134 18min 15min 3 121 20min 15min 3 Page 51 of 58 ...
Page 53: ...Page 52 of 58 ...