5.3.3 USER-DEFINNED PROGRAM
Fig.5.3-33 Fig.5.3-34 Fig.5.3-35
Click
“
User-definned
”
(Fig.5.3-33) enter into interface Fig.5.3-34.Press
“
”
to change the number of
“
Temperature/
Vacuum times/Sterilization time/Drying time
”
.
Temperature range:121
℃
or 134
℃
. Vacuum times range: 1 or 3. Sterilization time range:4~60. Drying time:1~60.
Then click
“
”
to save and exit,enter into Fig.5.3-35 interface.Then press
“
Start
”
buttom to start.Refer to
“
chapter
”
5.3.1
or 5.3.2 PRE-VACUUM PROGRAM
to check the running state diagram.
WARNNING:
If the result of sterilization is
“
FAILURE
”
, medical
instruments of this time can not be used. And it's need to contact
professional engineer to check the machine.