5.3.2 ONE TIMES PRE-VACUUM PROGRAM
(
SOLID
PLASTIC
)
Fig.5.3-19 Fig.5.3-20 Fig.5.3-21 Fig.5.3-22
C
lick
“
start
”
botton of Fig.5.3-19 interface,the sterilization program starts running, shows Fig.5.3-20 or Fig.5.3-21 interface.
When temperature of heating coil(T3) rised and less than 100
℃
,screen will show Fig.5.3-20 interface (the heating coil pre-heating stage).
When temperature of heating coil(T3) rised and over than 100
℃
,screen will show Fig.5.3-21 interface (the first times pre-vacuum stage).
Click
“
”
to check T3 temperature.(Fig.5.3-22)
5.3.2.1.THE FIRST TIMES PRE-VACUUM STAGE(Fig.5.3-21)
when the T3 temperature rised over than 100
℃
,the sterilizer will start evacuating and the control system will open the vacuum pump
solenoid valve(Ev4).LCD show
“
00:03:59
”
time countdown.If the vacuum value reaches Pv value ahead of time, sterilization
program goes into heating stage in advance.
ATTENTION:If the vacuum value of less than Pv (factory default setting Pv=-0.800bar) in
“
00:03:59
”
, there will be with alarm E12.
5.3.2.2.THE FIRST TIMES HEATING STAGE(Fig.5.3-23)
When the vacuum value reachs Pv value,the contrl system will open the water pump-solenoid valve(Ev2).The steam generator start inject the
steam into chamber,the temperatures(T1 and T2) and pressure inside the chamber will be rised at the same time.
5.3.2.3.THE FIRST TIMES EXHAUST STAGE(Fig.5.3-24)
When the pressure inside the chamber reachs 1.100bar,the control system will open the exhaust-solenoid valve(Ev1) for exhausting untill pressure
downs to P=0.000bar.LCD shows
“
00:01:29
”
time countdown.
5.3.2.4.THE SECOND TIMES HEATING STAGE(Fig.5.3-25)
When the pressure inside the chamber reaches 0.000bar,the control system will open the water pump-solenoid valve(Ev2).The steam generator
start to inject the steam into chamber,the temperature(T1 and T2) and pressure inside the chamber will be rised at the same time.
5.3.2.5.THE START STERILIZATION STAGE(Fig.5.3-26)
When the temperature inside chamber reaches
,the control system will start sterilizing.LCD shows
“
sterilization time
”
countdown.
134
℃
(
) or 121
℃
(
)
SOLID
PLASTIC
Page 33 of 58
Summary of Contents for JN-23
Page 1: ...EURO CERT 1128 0197 English JN 23 MODEL SteamSterilizer InstructionManual CLASS B and S ...
Page 4: ...Page 03 of 58 ...
Page 5: ...Page 04 of 58 ...
Page 7: ...Page 06 of 58 ...
Page 8: ...Page 07 of 58 ...
Page 10: ...Page 09 of 58 JN 23 overall space required ...
Page 12: ...Page 11 of 58 ...
Page 13: ...Page 12 of 58 ...
Page 14: ...Page 13 of 58 ...
Page 40: ...Page 39 of 58 ...
Page 41: ...Fig 6 2 1 Fig 6 2 2 Fig 6 2 3 Page 40 of 58 ...
Page 42: ...Fig 6 2 4 Fig 6 2 5 Fig 6 2 6 Chapter 4 3 2 Chapter 4 3 2 Page 41 of 58 ...
Page 43: ...6 2 1 Periodic Maintenance Page 42 of 58 ...
Page 44: ...6 3 Extraordinary Maintenance Page 43 of 58 ...
Page 51: ...Page 50 of 58 ...
Page 52: ...134 04min 15min 3 134 18min 15min 3 121 20min 15min 3 Page 51 of 58 ...
Page 53: ...Page 52 of 58 ...