Alarms, displays, and views during monitoring 59
7.6.1 Information signals from the alarm unit next to
the display
After the monitor is switched on, an acoustic reminder signal is
emitted every twenty seconds until plausible data have been detected.
7.6.2 Information
signals
from the sound aperture
between the sockets
A pulsating tone is emitted if the external power adapter is discon-
nected and no batteries are installed.
7.7
The visual alarm signals
A high-priority alarm, i. e. physiological alarm, causes
the alarm LED to flash red.
A medium-priority alarm, i. e. technical alarm, causes
the alarm LED to flash yellow.
7.8
Status line displays
During monitoring the status line is displayed in all views.
Fig. 40 The status line displayed in all views
The monitor’s text messages appearing on the left are explained in
detail in the section “Alarm messages – meanings and other informa-
tion” on page 62. On the right of the status line are three symbols.
Power supply