Intended
use 17
2.5
Limitations on VitaGuard®’s intended use
Even when operated in accordance with its intended use, VitaGuard®
cannot detect all life-threatening situations under certain unfavor-
able conditions.
The monitoring of SpO
2
and pulse rate is adversely affected when the
patient moves vigorously or is vigorously moved.
When the sensor is not attached correctly, ambient light can falsify
measurements. One remedy is to cover the sensor with a dark or
opaque material.
The monitor operates properly only when the SpO
2
sensor is correctly
attached.
2.6
Information for the doctor on these
operating instructions
In full knowledge of these operating instructions, the treating doc-
tor must decide:
whether the caregivers have to be trained in the performance of
resuscitation measures,
how the caregivers can be best prepared for monitoring and above
all for the measures that must be taken in the event of an alarm,
which view should be displayed
Information on
Settings protection
that sets the display modes and
user configurations can be found on page 72.
“Information for the doctor and qualified medical staff” is found on
page 78.