Information for the doctor and qualified medical staff 99
current value falls below the average by
more than the value set here and when
SpO
2
alarms
is set to
Limits & trends
,
an alarm is triggered.
SpO
2
alarms ..........................
When
SpO
2
alarms
is set to
Limits
only
, alarms are reported only when
the measured values violate the set
alarm limits. When
SpO
2
alarms
is set
to
Limits & trends,
alarms are reported
when the measured values violate the
set alarm limits AND when they deviate
from the average
SpO
2
measured over
the set interval.
9.11 Settings in the
Pulse rate
menu
(Settings
protection Off)
For this setting, Settings protec-
tion must be set to Off as ex-
plained under “System\ Settings
protection On, Limited, Off” on
page 72.
The possible settings when
Settings protection
is set to
Limited
are explained in the
section “
Pulse rate
menu –
alarm settings (Settings
protec-
tion Limited)
” on page 76.
Factory settings are shown in
bold type.
Fig. 77 Settings in the Pulse rate menu