GK-7PLUS CONTROLLER INSTRUCTION MANUAL
11
3.2.2. SUPERVISOR mode / REGULARIZATION
This option allows adjusting the level shown by the kit to the real level that there is in the tank. Because of
some differences in the accuracy of some instruments, which are used to count the entered and supplied
liquid, it is possible that there can be some imbalances between the shown value by the kit and the
measured value. This option allows making a "regularization" of this level.
Once selected one of the available tanks, the display will show:
REGULARIZ. LIQUID X
CURRENT QUANTITY:
The kit will wait that the current litre (or the volume unit with which is working) number is typed. Once the
quantity is typed and confirmed with the key <OK>, the kit will have this value as the current tank level.
3.2.3. SUPERVISOR mode / CALIBRATION
This option allows the meter calibration. The kit has built in an automatic electronic calibration process by
means of passing a known amount of liquid through the meter; while the kit "counts" the pulses generated
by the meter. Once the liquid has passed, it will be indicated to the kit the transferred amount with which the
kit can calculate the relation that there is between the pulses and the volume. Independently of this system,
the kit has also built in the traditional system of the whole factor, as it will be described below.
Procedure:
Accepting the CALIBRATION option, after putting the pointer and pressing the key <OK>, the kit
will start the pump (independently of the nozzle position) and will show the following message on the
display:
SUPPLY KNOWN AMOUNT
AND PRESS OK
In this point, fill a recipient with a known volume. The best option is to use a homologated dekalitre. Once
filled, press <OK>, and the kit will stop the pump and show the following message:
TYPE SUPPLIED
AMOUNT:
where the supplied volume must be typed. Remember that there is the possibility to type in decimals.
Consult the section 1. General Instructions to use the keypad.
Once typed this amount and confirmed with the key <OK>, the display will show the following message:
NEW FACTOR FFFFFF
OK = ACCEPT
EXIT = REJECT
where FFFFFF is the new calculated factor of calibration and represents the number of hundred thousandth
of litres (or the volume unit with which is working) that the count will be increased by each received pulse,
having the option to accept the new value through the key <OK> or reject it through the key <EXIT>.
If the kit has not noticed any pulse or the typed quantity has been equal to 0, it will not be possible to make
the calculation and the shown message will be: