Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Ergostik
guaranteed by the manufacturer and for the safety of the patient
and operator.
Storage / Transport:
min.
max.
Temperature:
-10 °C
+50 °C
Relative air humidity:
0 %
95 %
non-condensing
Operation:
min.
max.
Environmental temperature:
+10 °C
+35 °C
Avoid extreme
fluctuations in
temperature!
Relative air humidity:
20 %
95 %
non-condensing
(at least 30 % for synthetic flooring)
Atmospheric pressure:
800 hPa
1080 hPa
Setup:
mobile
Space requirement:
(complete with medical cart)
(L) 900 mm x (W) 900 mm x (H) 1900 mm
Required floor load capacity: 150 kg/m²
Flooring:
stable; even; conductive
preferably no synthetic material
Environment / room:
Closed, clinical area
air-conditioned
not an explosive or flammable environment!
not near an MRI!
well ventilated
Operating Mode:
Continuous operation