Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Page 46
Version: 7 | Release date: 02. August 2021
Attention
Ergostik could be damaged.
Reason: Electrostatic discharges. Therefore:
•
Preferably no floor made of synthetic material!
•
Otherwise a relative air humidity of at least 30 % is
required!
4.1.3.1
Data Connection
All measurement results and graphical data can be displayed on
the screen as well as printed out via the Windows printer
interface. Any printer compatible with Windows can be used for
this.
4.1.3.2
Equipment Cart with Isolating Transformer
The cart, approved by the manufacturer, meets the
requirements of the IEC 60601 series of standards, so its use is
recommended.
You will find more detailed information in:
•
Chap. 2.2.1 "Original Spare Parts / Accessories /
optional Expansions"
•
Respectively consult with your authorised specialist
retail partner.
The hardware connection between the device and computer is
established via an integrated USB interface and USB cable.
This is considered to be a spare part approved by the
manufacturer and can be obtained from an authorised
specialist retail partner.
•
See also chap. 2.2.1 "Original Spare Parts /
Accessories / optional Expansions"