No flow signal
Check that the Ergoflow is correctly positioned in
the Ergoclip
Check the tube connection between Ergoclip
and Ergostik
Drift on flow signal
Check that the Ergoflow is correctly positioned in
the Ergoclip
Check the tube connection between Ergoclip
and Ergostik
Check for moisture in the tube / connection
piece of the Ergoflow
Perform volume calibration
No gas signal
Check that the Ergoflow is correctly positioned in
the Ergoclip
Check the tube connection between Ergoclip
and Ergostik
Check for moisture in the tube / connection
piece of the Ergoflow
Perform gas calibration
Drift on gas signal
Note the warm-up time of the Ergostik
(15 minutes)
Check for moisture in the tube / connection
piece of the Ergoflow
Perform gas calibration
Software error message
"Unit warm-up not
completed.
Note the warm-up time of the Ergostik
(15 minutes)
Software error message
"Flow out of measuring
range".
Check that the Ergoflow is correctly positioned in
the Ergoclip
Check tube connection between Ergoclip and
Ergostik
Check for moisture in the tube / connection
piece of the Ergoflow