Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Page 60
Version: 7 | Release date: 02. August 2021
7.5
Calibrating Ergostik
There are various processes to be carried out for Ergostik in
order to calibrate or validate the volume and gas measurement.
The required intervals can be found in the tables below.
For details on individual calibrations, please read:
•
The separate IFU: Calibration
Interval
Calibration
Once per day
Gas calibration
Once per day, and after changing the flow
sensor
Volume calibration
Once per week
Flow linearity test
After Maintenance / Cleaning
Volume calibration
7.6
Using Ergostik / Performing Measurements
Descriptions of the individual examinations can be found in the
separate IFU:
•
CPET
And for general use of the software:
•
BLUE CHERRY®