![GERATHERM 820658 Instructions For Use Manual Download Page 44](http://html1.mh-extra.com/html/geratherm/820658/820658_instructions-for-use-manual_2216237044.webp)
Instructions for Use
Spirostik
Serial no.: xx|8|001|yyy and 2001xxxxx
Page 44
Version: 6 | Release date: 14. July 2021
7
Operating Instructions
Attention
Spirostik could be damaged.
Reason: Penetrating liquid. Uncleanliness. External exposure.
Therefore:
•
Do not use any liquids near the Spirostik!
•
Do not expose Spirostik or the entire system to dust or
other contamination!
•
Do not drop any objects on Spirostik! Do not lay any
objects on it! Never push foreign objects into the
housing!
7.1
Checking for Worn Parts
The Spirostik should be checked for defective wearing parts
each day before beginning treatments.
•
You can get a description in chap. 8 “Servicing /
Maintenance”
And:
Possible danger to life.
Reason: Misdiagnosis caused by measurement error
due to incorrect components or improper use.
Therefore:
•
Replace the flow sensor and the mouthpiece in
case of a functional error!
•
Always follow instructions (see chap. 8 "Servicing
/ Maintenance" and chap. 9 "Cleaning and
Disinfection")!