![GERATHERM 820658 Instructions For Use Manual Download Page 31](http://html1.mh-extra.com/html/geratherm/820658/820658_instructions-for-use-manual_2216237031.webp)
Instructions for Use
Spirostik
Serial no.: xx|8|001|yyy and 2001xxxxx
Version: 6 | Release date: 14. July 2021
Page 31
3.3
Operation / Servicing and Maintenance
Danger to life.
Reason: Electric shock. Explosion. Device damage.
Unpredictable movement of metal parts. For this:
Comply with the required installation and operating
conditions! Therefore:
•
In principle, observe chap. 12 “Technical
Specifications”!
•
Do not operate Spirostik if there are flammable or
explosive gases in the room!
•
Do not operate Spirostik near the magnetic field
of an MRT system!
Possible danger to life.
Reason: Electric shock. Cross-contamination.
Misdiagnosis caused by measurement error. Device
damage. Therefore:
•
Ensure that the used components are
undamaged and that you are working in a careful
way!
•
Do not use consumable items with a limited
expected service life after their use-by date has
been reached!
•
Prior to each use, visually inspect the total
system (housing, connectors, etc.) for any
damages!
•
If there are any damages, do not operate the
system. The damaged parts must be replaced or
repaired properly!
•
Calibrate Spirostik at the intervals stated!